Proceq 34310000 Operating Instructions Manual Download Page 5

© 2017 Proceq SA 

9

8  

               © 2017 Proceq SA

2.3  The RockSchmidt Display

Single shot   Statistics

Review

 Conv. Curve

Units

Correction factor

A typical display after an impact shows:

                      

A pseudo-analogue scale mimicking the mechanical drag pointer.
The actual rebound value with a two and a half digit number.
A counter displaying the actual number of valid impacts in a series of measurements.

If there is a battery or low impact energy alert (see chapter 8), it is displayed between the unit symbol 
“R” and the rebound value.

2.4  Philosophy of the User-Interface

These simple actions are all that is required to navigate around the menu, set up the RockSchmidt, 
perform measurements and review results and settings.

TILT

Tilting means that you raise the left side (plunger side) of the device to scroll to the 
icons to the left of the “centred” icon or you raise the right side of the device (SELECT 
button side) to scroll to the icons to the right of the “centred” icon.

ROLL

Rolling means that the device remains horizontal, but the display is rolled away from 
you or towards you.

SELECT

Press the SELECT button.

TIP

Depress the plunger a short distance, but not enough to trigger it.

IMPACT

Depress the plunger until it triggers.

2.5  Sample Preparation

When performing rebound measurements on core samples, both ASTM and ISRM recommend the 
use of a heavy specimen holder to prevent damage to the cores on impact. Ideally it should have 
a semi-circular groove with the same diameter as the core, but it may also be a V-shaped groove.

Please refer to the guidelines listed in chapter 2.1 for more details of both specimen preparation and 
suitable specimen holders.

When carrying out field testing it may be necessary to smooth the test surface using the grinding 
stone provide. Please note however, that this may affect the test result. Further information about 
this is contained in a research paper of the University of Oxford via the Proceq website  “The use of 
the Schmidt Hammer and Equotip for rock hardness assessment in geomorphology and heritage 
science: a comparative analysis”.

2.6  Correct use of the Hammer

Perform a few test impacts with the RockSchmidt before taking any measurements which you are 
going to evaluate. The single shot mode is useful for this purpose. Perform a consistency check on 
the test anvil.

The number of test impacts required for a valid rebound 
number is defined in the guidelines. Unless you are using 
the weathering grade mode, this will typically be at least 10 
impacts. The individual impacts should be separated by at 
least the width of the plunger. Ensure that the hammer is per-
pendicular to the test surface. Push the hammer against the 
surface at moderate speed until the impact is triggered.

NOTE! The plunger generates a recoil when it deploys. Always hold the RockSchmidt 
hammer in both hands, perpendicular to the test surface, before you trigger the impact!

Summary of Contents for 34310000

Page 1: ...Made in Switzerland ROCK SCHMIDT OPERATING INSTRUCTIONS...

Page 2: ...terface 8 2 5 Sample Preparation 9 2 6 Correct use of the Hammer 9 3 Optional Settings 10 3 1 Units 10 3 2 Statistics 10 3 3 Conversion Curve 11 3 4 Correction Factor 11 3 5 Checking the Current Setti...

Page 3: ...ration and maintenance of the instrument and its components Unauthorized structural modifications to the instrument and its components Serious damage resulting from the effects of foreign bodies accid...

Page 4: ...hut down automatically after 5 minutes of inactivity Standards and guidelines applied The following standards and guidelines have been applied in RockSchmidt for the procedure to determine the rebound...

Page 5: ...en holder to prevent damage to the cores on impact Ideally it should have a semi circular groove with the same diameter as the core but it may also be a V shaped groove Please refer to the guidelines...

Page 6: ...nd the range Once downloaded to a PC RockLink also calculates the median and the mode If a conversion curve has been selected tilting the instrument shows the UCS in the chosen unit Tilting again disp...

Page 7: ...are not saved Select single shot mode Confirm or cancel the action 5 2 Measurement Series The number of impacts required for a valid rebound number depends on the statistics method selected Having sel...

Page 8: ...lect review list Scroll in the list by ROLLING the hammer away from you or towards you The arrow indicates the current series Select summary for details of the series Select to delete the series Selec...

Page 9: ...over time The RockSchmidt checks for correct impact energy with each and every impact see above Nevertheless it is recommended to check the calibration on a regular basis Set the hammer to single shot...

Page 10: ...ton to establish the connection 9 1 Viewing the Data The data stored on your RockSchmidt will be displayed on the screen The test series is identified by the Impact counter value There is a Name colum...

Page 11: ...exported in the format defined by the active Tab i e either in series or summary format To open the file in Excel locate the file right click on it and Open with Microsoft Excel The data will be opene...

Page 12: ...eters and the hammer type RockSchmidt N for which it is applicable NOTE The curve may be exponential or a first second or third order polynomial ISRM recommend the use of exponential curves for UCS an...

Page 13: ...07 info asia proceq com Proceq Rus LLC Ul Optikov 4 Korp 2 lit A Office 410 197374 St Petersburg Russia Phone Fax 7 812 448 3500 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharja...

Reviews: