9.7  Custom Statistics

RockLink allows you to create a user defined statistics method and to download this onto the 
RockSchmidt.
Select the menu item File – Application settings” “Device – Custom statistics…” to bring up the 
following window.

Define the statistics method as desired and load this setting onto the RockSchmidt by clicking on 
the arrow in the center of the window.

See chapter 3.2 for selecting custom statistics on RockSchmidt.

9.8  Custom Curves

RockLink allows you to create custom curves specific to the type of rock under test or to use pre-
viously defined UCS and E-modulus correlation curves and upload them onto the RockSchmidt.

Select the menu item “Device – Custom curves….”

This will bring up the following window.

Here you can review custom curves you have 
previously created, create new curves, delete 
curves and download curves onto the Rock-
Schmidt. 
Click on “Create new curve…” to bring up the 
following window.

Type in the name for the new curve and enter the parameters and the hammer type (RockSchmidt 
N) for which it is applicable.

NOTE! The curve may be exponential or a first, second or third order polynomial.

ISRM recommend the use of exponential curves for UCS and modulus of elasticity

.

 If 

it is modulus of elasticity curve, be sure to tick the box below the hammer type. When 

using this type of curve, it is necessary to select GPa as the unit on RockSchmidt (See 

chapter 3.1).

It is also possible to limit the valid range for the R-reading. Values outside this range will not be 
evaluated by the RockSchmidt and this will be indicated on the instrument and also in downloaded 
data in RockLink.
Once you have entered the data, the curve will be displayed.

Click on “Create” to complete the action.

Uploading the curve onto the RockSchmidt

From the “Upload customer curves…” window, 

Select the curve you have just created and click on “Save on RockSchmidt…”.

You can store up to 3 custom curves at a time on the RockSchmidt. Choose the location of the curve 
(C1, C2, or C3) and upload. A warning window is displayed. Click on “Yes” to complete the action.

NOTE! You may create up to 200 curves which will be available on RockLink.

See chapter 3.3 for selecting the conversion curve on the RockSchmidt.

Summary of Contents for 34310000

Page 1: ...Made in Switzerland ROCK SCHMIDT OPERATING INSTRUCTIONS...

Page 2: ...terface 8 2 5 Sample Preparation 9 2 6 Correct use of the Hammer 9 3 Optional Settings 10 3 1 Units 10 3 2 Statistics 10 3 3 Conversion Curve 11 3 4 Correction Factor 11 3 5 Checking the Current Setti...

Page 3: ...ration and maintenance of the instrument and its components Unauthorized structural modifications to the instrument and its components Serious damage resulting from the effects of foreign bodies accid...

Page 4: ...hut down automatically after 5 minutes of inactivity Standards and guidelines applied The following standards and guidelines have been applied in RockSchmidt for the procedure to determine the rebound...

Page 5: ...en holder to prevent damage to the cores on impact Ideally it should have a semi circular groove with the same diameter as the core but it may also be a V shaped groove Please refer to the guidelines...

Page 6: ...nd the range Once downloaded to a PC RockLink also calculates the median and the mode If a conversion curve has been selected tilting the instrument shows the UCS in the chosen unit Tilting again disp...

Page 7: ...are not saved Select single shot mode Confirm or cancel the action 5 2 Measurement Series The number of impacts required for a valid rebound number depends on the statistics method selected Having sel...

Page 8: ...lect review list Scroll in the list by ROLLING the hammer away from you or towards you The arrow indicates the current series Select summary for details of the series Select to delete the series Selec...

Page 9: ...over time The RockSchmidt checks for correct impact energy with each and every impact see above Nevertheless it is recommended to check the calibration on a regular basis Set the hammer to single shot...

Page 10: ...ton to establish the connection 9 1 Viewing the Data The data stored on your RockSchmidt will be displayed on the screen The test series is identified by the Impact counter value There is a Name colum...

Page 11: ...exported in the format defined by the active Tab i e either in series or summary format To open the file in Excel locate the file right click on it and Open with Microsoft Excel The data will be opene...

Page 12: ...eters and the hammer type RockSchmidt N for which it is applicable NOTE The curve may be exponential or a first second or third order polynomial ISRM recommend the use of exponential curves for UCS an...

Page 13: ...07 info asia proceq com Proceq Rus LLC Ul Optikov 4 Korp 2 lit A Office 410 197374 St Petersburg Russia Phone Fax 7 812 448 3500 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharja...

Reviews: