OPPERATION INSTRUCTIONS
1. Connecting the System
1.1 Connect the power cord into a
properly grounded electrical outlet with
a sufficient amperage rating.
1.2 VOLTAGE: 120 Volt – The GS-601
requires a full 20 amp grounded circuit
for 120 volt operation. 240 Volt –The
GS-601 requires a full 10 amp grounded
circuit for 240 volt operation.
1.3 EXTENSION CORDS If used, should
be as short as possible and not less than
12 gauge.Heavy duty cords are
recommended.
1.4 CIRCUITS that have less than 15
amps or that have other high demand
equipment or appliances(especially more
than one heat seal machine) plugged in,
should not be used.
CAUTION: Failure to follow these
instructions will cause:
1. Erratic controller functions.
2. Inaccurate displays & slow heat-up.
3. The circuit breaker to disengage.
NOTE: If the supply cord is damaged, it
must be replaced by the manufacturer,
its service agent or a similarly qualified
person in order to avoid hazard.
2.1 Switch The System On
2. Turn On The System
3.1 Press the “set” button once time.
“Temperature” and “Setting” lights will
illuminate, press the “up” and “down”
button to set target start temperature.
3. Adjusting The Temperature
4. Adjusting The Time
4.1 Once you have adjusted the
temperature, press the “set” button again.
“Time” lights located in the display will
illuminate. (See Image On Next Page)