PSS500 CONTROL BOARD JUMPERS - FIELD CONFIGURATION
(CONTINUED)
Step 7
Configure the board by setting the appropriate jumpers. A good
reference is
DYK #202 PSS5 Control Board Jumper
Configuration,
which will help in understanding how to check that
the jumpers are configured to match the Opto-22 relays.
If this is a new board that will be set up identically as the old
board, then disassemble the old board as described above. Copy
the jumper positions from the old to the new board.
If this is a new board and no old board is available, the interface
board jumpers must be set using the electrical diagram provided
in the Operator Manual and in the electrical box shipped with the
sterilizer. Without this diagram, it will be extremely difficult, if not
impossible, to correctly configure the board. If needed, the
electrical diagram can be requested from PRIMUS.
Be cautious when setting and removing the jumpers to avoid
bending the jumper pins.
Once the jumpers are set, it is a good idea to document their
placement with another photograph before re-assembling the two
boards.
If the four stand-offs were taken apart, re-assemble them before
proceeding.
Step 8
Carefully orient the SBC196 board with the bottom pins matching
the sockets where they will be connected. There is a short and
long socket to help orient the board.
Step 9
Once the SBC196 board is oriented properly and the stand-offs
are matched with the hole in the corner of the board, gently press
down to engage the pins and the mounting sockets. Do not press
down firmly before ensuring the pins and sockets are properly
matched.