Presys T-1200PIR Technical Manual Download Page 4

PRESYS

 Instruments 

T-1200PIR

Page 1 

1.0 - Introduction 

T-1200PIR

The Calibrator for Infrared Thermometer T-1200PIR generates temperature 

values in order to enable the calibration of infrared thermometers. The temperatures 
generated have high accuracy, in an area of uniform temperature and emissivity. 

This calibrator also has input for thermocouples and resistance thermometers, as 

well as current meter function. The T-1200PIR generates temperature from 50 to
1200 ºC. 

  Has input for thermocouples, RTDs and thermoswitches reading.  

  The use of external standard thermometer is not required*. 

  Carries out completely automatic calibrations with or without the use of a computer. 

  Accuracy of 3 ºC, 0.2 ºC stability and 0.1 ºC resolution. 

  Documenting capabilities: communication with computer and ISOPLAN Calibration 

Software. 

  Portable and compact, has strap for easy transport. 

* For higher performances, it is possible to make use of an external thermocouple sensor 
type "N" connected to its own input.   

It has extensive programming capabilities, where the calibrator is programmed to 

generate pre-set temperature. 

Compatible with automatic calibration and documenting software, with application 

of ISOPLAN software on PC / Windows platform, using the serial port for connection 
between the PC and the calibrator that communicate via RS-232 or RS-485. ISOPLAN 
software can register thermometers and factory instruments, generate work orders, 
produce and print calibration certificates and reports, that is, all the information of power 
is brought into the environment of calibrations. 

presys

T-1200PIR

T-1200PIR

The Calibrator for Infrared Thermometer T-1200PIR generates temperature 

The Calibrator for Infrared Thermometer T-1200PIR generates temperature 

values in order to enable the calibration of infrared thermometers. The temperatures 

values in order to enable the calibration of infrared thermometers. The temperatures 
generated have high accuracy, in an area of uniform temperature and emissivity. 

generated have high accuracy, in an area of uniform temperature and emissivity. 

This calibrator also has input for thermocouples and resistance thermometers, as 

This calibrator also has input for thermocouples and resistance thermometers, as 

well as current meter function. The T-1200PIR generates temperature from 

well as current meter function. The T-1200PIR generates temperature from 

  Has input for thermocouples, RTDs and thermoswitches reading.  

  Has input for thermocouples, RTDs and thermoswitches reading.  
  The use of external standard thermometer is not required*. 

  The use of external standard thermometer is not required*. 
  Carries out completely automatic calibrations with or without the use of a computer. 

  Carries out completely automatic calibrations with or without the use of a computer. 

ºC

ºC, 0

, 0.2 

.2 ºC

ºC stability and 0.1 

 stability and 0.1 ºC

ºC resolution. 

 resolution. 

  Documenting capabilities: communication with computer and ISOPLAN Calibration 

  Documenting capabilities: communication with computer and ISOPLAN Calibration 

  Portable and compact, has strap for easy transport. 

  Portable and compact, has strap for easy transport. 

For higher performances, it is possible to make use of an external thermocouple sensor 

For higher performances, it is possible to make use of an external thermocouple sensor 

type "N" connected to its own input

type "N" connected to its own input..   

  

It has extensive programming capabilities, where the calibrator is programmed to 

It has extensive programming capabilities, where the calibrator is programmed to 

generate pre-set temperature. 

generate pre-set temperature. 

Compatible with automatic calibration and documenting software, with application 

Compatible with automatic calibration and documenting software, with application 

of ISOPLAN software on PC / Windows platform

of ISOPLAN software on PC / Windows platform
between the PC and the calibrator that communicate via RS-232 or RS-

between the PC and the calibrator that communicate via RS-232 or RS-
software can register thermometers and factory instruments, generate work orders, 

software can register thermometers and factory instruments, generate work orders, 
produce and print calibration certificates and reports, that is, all the information of power 

produce and print calibration certificates and reports, that is, all the information of power 
is brought into the environment of calibrations. 

is brought into the environment of calibrations. 

Summary of Contents for T-1200PIR

Page 1: ...momether T 1200PIR EM0277 00 R Technical Manual Q Y U T ALI ENT M S E Y G S A T N E A M M p r e s y s p r e s y s p r e s y s p r e s y s Infrared Thermomether p r e s y s Infrared Thermomether p r e s y s p r e s y s p r e s y s ...

Page 2: ...ters that protect not only p r e s y s The equipment is provided with electromagnetic interference filters that protect not only the mains but also the equipment itself against noise These filters have no function if p r e s y s the mains but also the equipment itself against noise These filters have no function if the unit is not earthed properly p r e s y s the unit is not earthed properly WARNI...

Page 3: ...ng the Strap for Transport 21 p r e s y s p r e s y s p r e s y s p r e s y s 3 p r e s y s 3 p r e s y s 4 p r e s y s 4 p r e s y s 4 p r e s y s 4 p r e s y s 6 p r e s y s 6 p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s 5 0 Calibrator Warning Messages p r e s y s 5 0 Calibrator Warning Messages p r e s y s 6 0 PID Control Parameters p r e s y s 6 0 PID Con...

Page 4: ... as current meter function The T 1200PIR generates temperature from p r e s y s well as current meter function The T 1200PIR generates temperature from Has input for thermocouples RTDs and thermoswitches reading p r e s y s Has input for thermocouples RTDs and thermoswitches reading The use of external standard thermometer is not required p r e s y s The use of external standard thermometer is not...

Page 5: ...ectric signal Independent circuitry for over temperature protection and safety p r e s y s Independent circuitry for over temperature protection and safety T p r e s y s T p r e s y s 1200PIR p r e s y s 1200PIR 50 to p r e s y s 50 to 1200 p r e s y s 1200 C p r e s y s C p r e s y s 110 or p r e s y s 110 or 220 p r e s y s 220 Va p r e s y s Vac p r e s y s c p r e s y s 50 60Hz p r e s y s 50 ...

Page 6: ...y s 2 5 ºC p r e s y s 5 0 ºF p r e s y s 5 0 ºF p r e s y s 1 5 ºC p r e s y s 1 5 ºC p r e s y s 3 0 ºF p r e s y s 3 0 ºF p r e s y s 1 0 ºC p r e s y s 1 0 ºC p r e s y s 2 0 ºF p r e s y s 2 0 ºF p r e s y s 0 7 ºC p r e s y s 0 7 ºC p r e s y s 1 4 ºF p r e s y s 1 4 ºF IEC p r e s y s IEC p r e s y s 60584 p r e s y s 60584 p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s...

Page 7: ...nted in the lower part of the pit of the p r e s y s Mounted in the lower part of the pit of the Mounted in the lower part of the pit of the p r e s y s Mounted in the lower part of the pit of the Metallic insert type black body cavity p r e s y s Metallic insert type black body cavity p r e s y s Must be introduced in the pit joined with p r e s y s Must be introduced in the pit joined with Must ...

Page 8: ...heath with the GND mV in order to prevent noise interference SCHEMATIC VIEW FOR MOUNTING OF THE BLACK BODY CAVITY p r e s y s Use a cable to short circuit the TC sensor sheath with the GND mV in order to p r e s y s Use a cable to short circuit the TC sensor sheath with the GND mV in order to p r e s y s SCHEMATIC VIEW FOR MOUNTING OF p r e s y s SCHEMATIC VIEW FOR MOUNTING OF THE BLACK BODY CAVIT...

Page 9: ...esides the thermo element readings are stored p r e s y s schedule and the calibration results besides the thermo element readings are stored in the calibrator memory For pyrometers with local indication the reading is entered p r e s y s in the calibrator memory For pyrometers with local indication the reading is entered manually and for pyrometers with response in mA mV or thermocouple reading i...

Page 10: ...e s y s selects thermocouple input Choose among the B p r e s y s B p r e s y s J p r e s y s J p r e s y s K p r e s y s K p r e s y s N p r e s y s N p r e s y s R p r e s y s R and p r e s y s and S p r e s y s S types p r e s y s types In the next menu the internal or manual cold junction compensation is chosen p r e s y s In the next menu the internal or manual cold junction compensation is c...

Page 11: ...PRESYS Instruments T 1200PIR Page 8 2 1 Input Connections Diagram p r e s y s p r e s y s ...

Page 12: ... is turned off or on respectively p r e s y s cavity is turned off or on respectively Battery state p r e s y s Battery state Display p r e s y s Display p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s normal p r e s y s normal p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s low p r e s y s low LOW BATTERY p r e s y s LOW BATTERY p r e s ...

Page 13: ...l p r e s y s The default value depends on the input signal p r e s y s DEFAULT p r e s y s DEFAULT p r e s y s p r e s y s p r e s y s p r e s y s 1 p r e s y s 1 2 p r e s y s 2 3 p r e s y s 3 4 p r e s y s 4 p r e s y s p r e s y s Selects the number of decimals of the cavity temperature and p r e s y s Selects the number of decimals of the cavity temperature and the set point value The defaul...

Page 14: ...ature and the transmitter reading in the same unit Thus p r e s y s displays the current temperature and the transmitter reading in the same unit Thus the error can be verified directly in ºC or ºF Select the option p r e s y s the error can be verified directly in ºC or ºF Select the option SC p r e s y s SC and press p r e s y s and press ENTER p r e s y s ENTER If p r e s y s If SELECT p r e s ...

Page 15: ... Choose one of the positions p r e s y s 1 to 8 representing each one of the memory positions Choose one of the positions WRITE p r e s y s WRITE can be made in an already used p r e s y s can be made in an already used memory position The calibrator confirms the overwriting confirmation with the p r e s y s memory position The calibrator confirms the overwriting confirmation with the CLEAR p r e ...

Page 16: ...tly with no need to enter the set point p r e s y s are used directly with no need to enter the set point beside the cavity temperature set point value in the p r e s y s beside the cavity temperature set point value in the programmable mode The number of the program is indicated by p r e s y s programmable mode The number of the program is indicated by n p r e s y s n Using the keys p r e s y s U...

Page 17: ...p r e s y s TAGMAN p r e s y s p r e s y s p r e s y s Before start programming configure the signal that will be read by the calibrator p r e s y s Before start programming configure the signal that will be read by the calibrator To calibrate pyrometers with local indication for example there is no p r e s y s To calibrate pyrometers with local indication for example there is no IN p r e s y s IN...

Page 18: ...ption p r e s y s to be calibrated from the tag list The automatic calibration starts by the option EXEC p r e s y s EXEC and the calibration data can be verified in option p r e s y s and the calibration data can be verified in option VERIF p r e s y s VERIF To calibrate p r e s y s To calibrate automatically not using the ISOPLAN there is a reserved place for one manual tag p r e s y s automatic...

Page 19: ...t the observed results are p r e s y s These data are p r e s y s These data are easily obtained in the thermometer user manual and this observation is needed to p r e s y s easily obtained in the thermometer user manual and this observation is needed to To perform the readings align the pyrometer with the black body cavity p r e s y s To perform the readings align the pyrometer with the black bod...

Page 20: ... s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s 6 0 PID Control Parameters p r e s y s 6 0 PID Control Parameters The T 1200PIR calibrators have a PID control algorithm to calculate the p r e s y s The T 1200PIR calibrators have a PID control algorithm to calculate the temperature control output p r e s y s temperature control output The black body cavity stability and r...

Page 21: ...essary to return the instrument to the factory p r e s y s Otherwise it may be necessary to return the instrument to the factory ENTER p r e s y s ENTER key You should p r e s y s key You should 9875 to access the calibration menu p r e s y s 9875 to access the calibration menu The password functions as a protection to calibration ranges After the p r e s y s The password functions as a protection...

Page 22: ... s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s 0 000 mV p r e s y s 0 000 mV 120 000 mV p r e s y s 120 000 mV p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s 0 000 mV p r e s y s 0 000 mV 600 000 mV p r e s y s 600 000 mV p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s p r e s y s 600 000 mV p r e s y s 600 000 mV 2400 000 mV p r e s y s 240...

Page 23: ...ring values according to the table below p r e s y s initial storing values according to the table below Go to the manual operating mode menu p r e s y s Go to the manual operating mode menu EXEC p r e s y s EXEC and generate all the seven p r e s y s and generate all the seven levels of temperature set points from the table writing down the value indicated by the p r e s y s levels of temperature...

Page 24: ... The straps for transport are included together with the calibrator It is very useful p r e s y s The straps for transport are included together with the calibrator It is very useful during the transport period and its fitting is very simple Just manually screw the screws p r e s y s during the transport period and its fitting is very simple Just manually screw the screws on each tip of the strap ...

Page 25: ... Instruments Inc Presys Presys www presyscorp com p r e s y s ...

Reviews: