PreSens OXY-1 ST Trace Instruction Manual Download Page 10

10 

 

 
 
 
 

© 2017 PreSens Precision Sensing GmbH                                                                   

OXY-1 ST (Trace) 

Set-Up 

 

 
 

4. 

Now start the PreSens Measurement Studio (PMS2) software on your PC / notebook. 
(The OXY-1 ST always has to be connected via USB before the software is started.) After 
successful initialization the software main screen is displayed and the OXY-1 ST will show 
in the 

Device

 section. 

 
For further details on how to use the PreSens Measurement Studio please refer to the 
software instruction manual. 
 
 

Important Considerations for USB Handling 

!

 

PreSens recommends the use of a dedicated USB 2.0 PCI Card to connect and 

handle USB PreSens Devices with a Desktop PC. 
 

!

 

In order to enhance the system stability avoid the use of USB hubs and connect 

PreSens Devices directly to your PC / notebook USB Ports. 
 

!

 

If possible, disconnect all other USB devices that are not in use, as they may reduce 

or disturb the USB resources of your PC / notebook. 
 

!

 

Docking stations may also reduce or disturb the USB resources of your PC / notebook 

and therefore affect the correct function of the software. 
 

!

 

It is also recommended to disable the Power Saving Settings of your USB Root 

controller. 

Fig. 3  

Connecting the temperature sensor 

Summary of Contents for OXY-1 ST Trace

Page 1: ...OXY 1 ST Trace TRANSMITTERS Instruction Manual ...

Page 2: ......

Page 3: ... registered trademarks of the respective owners The publisher and the author make no claim to these trademarks While every precaution has been taken in the preparation of this document the publisher and the author assume no responsibility for errors or omissions or for damages resulting from the use of information contained in this document or from the use of programs and source code that may acco...

Page 4: ......

Page 5: ...recision Sensing GmbH www PreSens de Table of Contents 1 Preface 7 2 Safety Notes 8 3 Set Up 9 4 Technical Data 11 5 Operational Notes 12 5 1 Optical Output 12 5 2 Maintenance 12 5 3 Service 12 6 Concluding Remarks 13 ...

Page 6: ......

Page 7: ...asuring system which allows a non invasive measurement They can be used in disposables Microsensors allow measurements with high spatial resolution whenever this is required Therefore they are ideally suited for the examination of small sample volumes for highly parallelized measurements in disposables and for biotechnological applications A set of different oxygen mini and microsensors flow throu...

Page 8: ... purpose PreSens is explicitly not liable for direct or indirect losses caused by the application of these measurement systems In particular it has to be considered that malfunctions can occur due to the naturally limited lifetime of the sensor depending on the respective application The set up of backup measurement stations is recommended when using the sensors in critical applications to avoid c...

Page 9: ...tion on sensor handling and calibration please refer to the respective sensor instruction manual www presens de manuals 2 Connect the USB cable to the connector on the OXY 1 ST and to an USB port of your PC notebook 3 If required you can now connect a Pt100 temperature sensor to the device There is a red mark on the temperature sensor connector of the OXY 1 ST as well as on the temperature sensor ...

Page 10: ...ns for USB Handling PreSens recommends the use of a dedicated USB 2 0 PCI Card to connect and handle USB PreSens Devices with a Desktop PC In order to enhance the system stability avoid the use of USB hubs and connect PreSens Devices directly to your PC notebook USB Ports If possible disconnect all other USB devices that are not in use as they may reduce or disturb the USB resources of your PC not...

Page 11: ...ware LED peak wavelength 633 nm TEMPERATURE SENSOR Potentiometric temperature sensor Pt100 Range 0 50 C Resolution 0 1 C POWER SUPPLY Supply voltage 5 VDC USB 2 0 Mini B Current Power 400 mA DIGITAL INTERFACE USB interface cable to PC Cable included ENVIRONMENTAL CONDITIONS Operating temperature 0 C to 50 C Storage temperature 20 C to 70 C Relative humidity 0 to 80 non condensing DIMENSIONS WEIGHT...

Page 12: ...ture entering the housing Never use benzine acetone alcohol or any other organic solvents The ST fiber connector of the sensor can be cleaned with lint free cloth or a cleaning implement for ST connectors only 5 3 Service Alignment rework or repair work may only be carried out by the manufacturer PreSens Precision Sensing GmbH Am BioPark 11 93053 Regensburg Germany Phone 49 941 94272100 Fax 49 941...

Page 13: ... you with an introduction to work with the OXY 1 ST fiber optic oxygen meter This manual does not claim to be complete We are endeavored to improve and supplement this version We are looking forward to your critical review and to any suggestions you may have You can find the latest version at www PreSens de With best regards Your PreSens Team ...

Page 14: ...cluding Remarks 2017 PreSens Precision Sensing GmbH www PreSens de Manufacturer PreSens Precision Sensing GmbH Am BioPark 11 93053 Regensburg Germany Phone 49 941 94272100 Fax 49 941 94272111 info PreSens de www PreSens de ...