6
U
se the “▲” or “▼” buttons to set the value between “-9.9” and “9.9”.
For example, the measured average temperature of the heating surface is 30
℃
, but the LED
temperature display shows 25
℃
. Use the
“▲” button to add 5
℃
(the difference in temperature
values). Then when you exit the settings selections (step 5) the LED temperature display will
read 30
℃
.
2)
Press the “SET” button again and “E” will be shown in the LED temperature display. Press
the “SET” button again and “00.5” will be shown.
DO NOT CHANGE THIS VALUE
. It is
a factory preset and changing it will cause the Slide Warmer to malfunction.
3)
Press the “SET” button again and “P” will be shown in the LED temperature display. Press
the “SET” button again and “15.0” will be shown. See step 6 for instructions on when to
adjust this value.
4)
Press the “SET” button again and “c” will be shown in the LED temperature display. Press
the “SET” button again and “10.0” will be shown.
DO NOT CHANGE THIS VALUE
. It is
a factory preset and changing it will cause the Slide Warmer to malfunction.
5)
Press the “SET” button again and your settings will be saved.
6)
If the Slide Warmer does not maintain its recalibrated temperature:
Press and hold the “SET” button for 3 seconds until “Sc” is shown. Then press the “SET” button
4 more times and “P” will be shown in the LED temperature display. Press the “SET” button
again to see the value of “P”; this is the temperature control value. This value is equal to P/10°.
This value can be adjusted so that the heating surface maintains a steady temperature. Use the
“▲” or “▼” buttons to set the value between “0.00” and “99.9”.
For example, if the LED temperature display value is “50.0” and the value of “P” is “10.0” the
heater will turn off and on at P/10 = 10/10 = 1°C. So the heater will turn off at 51°C and heat
back up if it falls below 49°C.
If the temperature rises too slowly or it cannot reach the “SET” temperature increase the value of
“P”. If the temperature rises too fast or it surpasses the “SET” temperature decrease the value of
“P”.
7)
Press the “SET” button again and the LED temperature display shows “c” (see step 4).
8)
Press the “SET” button again to save your settings.
9)
Monitor for 10 minutes and readjust values as necessary.