background image

 

5.1 Operation

 

 
 
1.

 

Plug the Slide Warmer into an electrical outlet and turn on the power switch on the side of 

the machine.  The LED temperature display will turn on letting you know the machine is on 

and is working. 

 

2.

 

Press the temperature “SET” button (#2) to see the current temperature setting. 

 

3.

 

Set  the  desired temperature (shown 

in  degrees  Celsius)  by  using  the  “▲”  or  “▼” 

(temperature control) buttons (#1).  The setting will appear in the LED temperature display 

(#3). 

   

4.

 

Once you have entered your desired temperature setting push the temperature “SET” button 

(#2).  This will

 

show the actual temperature of the heating surface. 

 

5.

 

Temperature range is from room temperature to 75

˚C 

±

 

2˚C. 

 

6.

 

Place slides flat on the Slide Warmer surface to help warm more efficiently.   

 
 

 

6.1 Troubleshooting 

 
 
Calibration testing: 
 

Turn the Slide Warmer on and allow it to warm up for 15 minutes. Then use a surface or infrared 
thermometer to measure the average temperature between the edge of the heating surface and its 
center. If the difference between the measured average temperature and the temperature shown 
on the LED temperature display is more than 5°C, the Slide Warmer needs to be recalibrated. 

 
How to recalibrate

 
1)

 

Adjust the difference between average surface temperature and displayed temperature:   

Press and hold the “SET” button for 3 seconds. The LED temperature display  will show “Sc”.  
Press the “SET” button again and you will see the recalibration value begin to flash in the LED 
temperature display. This value can be adjusted so that the LED temperature display value and 
the measured average temperature of the heating surface are the same (the default value is 0.0).  
 

Summary of Contents for XH-2004

Page 1: ...o obtain the best results of this equipment Warning To prevent fire or shock hazard do not expose this unit to rain or moisture Unless labeled otherwise this equipment should be used with AC 120V 60Hz...

Page 2: ...1 1 1 Table of Contents Section Table of Contents Page 1 Safety Page 2 Contents Specifications Page 3 Control Panel Setup Page 4 Operation Page 5 Troubleshooting Page 5 6 Warranty Page 7...

Page 3: ...ing the fuse 7 If a problem occurs with the device contact your dealer Individuals that are not qualified to repair this device should not attempt to do so the result could be additional damage that w...

Page 4: ...26 7 x 5 in 27 x 68 x 13 cm 9 x 11 5 x 4 5 in 23 x 29 x 11 cm Net weight 23 Ibs 8 Ibs Dimensions packaging 29 x 13 x 7 in 74 x 33 x 18 cm 14 x 11 x 7 in 36 x 28 x 18 cm Gross weight 23 10 Ibs Capacit...

Page 5: ...isplay C 4 Heat indicator lamp 4 2 Setup 1 Open box and remove slide warmer from packaging 2 Place slide warmer onto flat surface 3 Plug AC power cord firmly into back of unit connect plug to power su...

Page 6: ...elp warm more efficiently 6 1 Troubleshooting Calibration testing Turn the Slide Warmer on and allow it to warm up for 15 minutes Then use a surface or infrared thermometer to measure the average temp...

Page 7: ...lide Warmer to malfunction 5 Press the SET button again and your settings will be saved 6 If the Slide Warmer does not maintain its recalibrated temperature Press and hold the SET button for 3 seconds...

Reviews: