Fig. 6:
Measurement cable and Hirschmann cable socket
Manufacturing a measurement cable
You can find additional information regarding measurement cable type and conductor cross-sections in
the technical data.
1. Observe the following steps to ensure optimum signal quality, whereby ground loops, differences
in potential or EMC will influence the measuring signal.
2. Use a cable with braided shield.
3. Connect the supply earth directly with the protective earth for power supply pack.
4. Use a differential measuring input and separate signal earth and supply earth.
5. Ensure that the potential difference for surge protection between supply earth and housing is
≤ 6 V.
6. Assemble the Hirschmann cable socket.
Assembling the Hirschmann cable socket
1. Prepare the Hirschmann cable socket as shown in the "Measurement cable and Hirschmann ca-
ble socket" diagram.
2. Solder in the connection cable as prescribed in the terminal connection diagram.
3. Assemble the Hirschmann cable socket.
Installation