background image

4

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

SPECIFICATIONS 

High Flow

Low Flow

Model

PM5580: 

80/20 heliox

PM5570: 

70/30 heliox

PM5480: 

80/20 heliox

PM5470: 

70/30 heliox

Primary Outlet

Flow Range

15 - 120 lpm

3 - 30 lpm

With both supply pressures at 50 psi (3.4 bar)

 with 

BLEED 

Closed

Auxiliary Outlet

Flow Range

2 - 100 lpm

0 - 30 lpm

With both supply pressures at 50 psi (3.4 bar) with 

BLEED 

Open

Bleed Flow @ 

100% F

I

O

2

13 lpm or less

at 50 psi (3.4 bar)

3 lpm or less

at 50 psi (3.4 bar)

Maximum 

Combined

Flow (All Outlets)

≥ 

120 lpm

 30 lpm

Bypass Flow

(Loss of Helium or 

Oxygen supply)

85 lpm

45 lpm

Bypass Alarm 

Activation

50 psi 

(3.45 bar)

60 psi

(4.14 bar)

50 psi

(3.45 bar)

60 psi

(4.14 bar)

13-25 psi

16-24 psi

18-22 psi

16-24 psi

0.9-1.7 bar

1.1-1.65 bar 1.2-1.5 bar 1.1-1.65 bar

Alarm Reset:

When pressure differential is 6 psi (0.4 bar) 

or less.

Alarm Sound Level:

 to 80 db at 1 ft (0.3 m)

Oxygen Concentration 

Adjustment Range:

PM5480

PM5580

20 - 

100%

PM5470

PM5570

30 - 100%

Gas Supply Pressure:

30 - 75 psi (2.1 - 5.2 bar) 

Heliox & Oxygen 

within 10 psi (0.69 bar) of each other

Mixed Gas Stability:

±1% Oxygen

Connection Types:

DISS Type - Heliox & Oxygen Inlets & Outlets 

and / or NIST Type - Heliox & Oxygen Inlets

Note

: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).

Summary of Contents for HELIO2 PM5500 Series

Page 1: ...ctions Model No PM5400 Series shown PM5500 Series SAVE THESE INSTRUCTIONS CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 No...

Page 2: ...ixtures of 20 oxygen 80 helium and 30 to 100 for heliox tank mixtures of 30 oxygen 70 helium The Helium and Oxygen Blender is a restricted medical device intended for use by qualified and trained pers...

Page 3: ...hread psi Pounds Per Square Inch lpm Liters Per Minute SAFETY Information Warnings and Cautions DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious...

Page 4: ...ender when alarm is sounding DO NOT use oil in or around the Blender DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender DO NOT use near any type of flame or flammable exp...

Page 5: ...Flow Loss of Helium or Oxygen supply 85 lpm 45 lpm Bypass Alarm Activation 50 psi 3 45 bar 60 psi 4 14 bar 50 psi 3 45 bar 60 psi 4 14 bar 13 25 psi 16 24 psi 18 22 psi 16 24 psi 0 9 1 7 bar 1 1 1 65...

Page 6: ...Helium Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195 The alarmed Oxygen Monitor Analyzer should comply with ISO 7767 to meet CE requirement Dryness and Composition for inlet gase...

Page 7: ...6 HELIO2 Blender HELIUM Oxygen Diagrams CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc PM5400 and PM5500 Models A B F E D C H G...

Page 8: ...or OFF With the bleed in the ON position this outlet delivers accurate oxygen concentrations in the following flows Model Flow Range High Flow 2 100 lpm Low Flow 0 30 lpm D Auxiliary Bleed Collar The...

Page 9: ...e auxiliary flow outlet maintains the same flow capacity and FIO2 accuracy as the Primary Outlets with Bleed Valve not engaged When the bleed flow is activated some of the heliox mixture will vent to...

Page 10: ...damaged 1 Secure Blender to wall or pole mount bracket 2 Connect heliox and oxygen supply lines from HELIO2 Blender to heliox cylinder and oxygen supply 3 Connect oxygen flowmeter to Blender outlet 4...

Page 11: ...Oxide DO NOT clean with aromatic hydrocarbons 1 Disconnect all gas connections and equipment before cleaning 2 Clean exterior surfaces with a cloth dampened with mild detergent and water 3 Wipe dry wi...

Page 12: ...0 18 32 4 28 8 25 2 23 4 21 6 21 2 20 7 18 4 18 0 19 34 2 30 4 26 6 24 7 22 8 22 4 21 9 19 4 19 0 20 36 0 32 0 28 0 26 0 24 0 23 6 23 0 20 4 20 0 21 37 8 33 6 29 4 27 3 25 2 24 8 24 2 21 4 21 0 22 39...

Page 13: ...rned Goods Authorization RGA number contact Precision Medical Inc All returns must be packaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in...

Page 14: ...inated or heliox concentration incorrect 5 Downstream device causing back flow or restricted flow 6 Supply pressure imbalanced 1 Use auxiliary outlet engage bleed 2 Recalibrate alarmed Oxygen Monitor...

Page 15: ...ke oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and...

Page 16: ...odel s PM5470EN PM5480EN PM5570EN PM5580EN MDD Class IIb Classification criteria Clause 3 2 Rule 11 of Annex IX of MDD As delivered the object of the declaration described above is in conformity with...

Reviews: