Precision Medical EasyMate PM2200 User Manual Download Page 9

8.  While holding the Portable System in the 

fill position, pull the vent-to-Fill lever to the 

open position (

figure 1

). A hissing noise 

should be noticeable.

NOTE

PM2200 Top fill ONLy

, maintain a 

slight downward pressure on the 

Portable  System  with  one  hand 

to  keep  the  device  steady  and 

maintain proper filling position.

9. release the vent-to-Fill lever on the Portable System as soon as you 

notice a change in the sound of venting gas followed by a dense, white 

vapor coming from the reservoir cover.

 

NOTE:

 The maximum time to fill a ....

PM2200 & PM2202

30 seconds.

PM2201

15 seconds.

figure 1

10. 

Disengagement:

 

PM2200:

•Disconnect the Portable System from 

the reservoir by pulling in an upright 

motion. Always hold the device with 

at least one hand when attempting to 

disconnect it. (

figure 2

)

 

PM2201 / PM2202:

 

•Disconnect the Portable System 

from the reservoir by rotating it in counterclockwise direction until 

the device separates. The device may now be removed from the 

reservoir.

NOTE: 1

.It is common to hear a hissing sound after the Portable 

System has been filled. This is the relief valve venting 

excess gas pressure. Upon disengaging the Portable 

System  from  the  reservoir  it  is  common  to  see 

condensation on or near the fill connector. 

2.It is common to have a few small droplets of liquid

oxygen coming from the fill connector when disengaging 

the Portable from the reservoir.

figure 2

CAUTION

If the vent-to-Fill lever fails to close and the hissing continues, remove 
the Portable from the reservoir by depressing the release button on 
the reservoir unit. The Portable will stop venting in a few minutes. 
The Portable may require as much as 30 minutes to restore normal 
operation.

8

Summary of Contents for EasyMate PM2200

Page 1: ...ON SAVE THESE INSTRUCTIONS Federal USA law restricts this device to sale by or on the order of a physician Precision Medical Inc 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 61...

Page 2: ...y bag 12 cleaning 13 Replacement parts 13 Maintenance 13 Returns 13 TROUBLESHOOTING 14 DISPOSAL 15 NOTES 16 WARRANTY 17 DECLARATION OF CONFORMITY 18 Receiving inspection Remove product from package an...

Page 3: ...RMENTS WILL PRODUCE OXYGEN ENRICHED ATMOSPHERE Symbol for SINGLE USE Applies to Cannula only 2 Symbol for Warning low temperature freezing conditions Symbol for Warning High pressure oxidizing gas vig...

Page 4: ...ow adult single lumen oxygen nasal cannula Pulse Volume Accuracy Within 15 of the nominal bolus value at each breath rate Trigger Method Inspiratory effort negative pressure from patient inhalation Br...

Page 5: ...flames Use and store Reservoir and Portable at least five feet away from equipment such as furnaces water heaters and stoves that may contain open flames Keep equipment in a well ventilated area at a...

Page 6: ...it is recommended that the Portable Liquid Oxygen System be qualified on patients in the situations it will be used rest exercise sleep This device is designed to operate with a single lumen adult na...

Page 7: ...g the various pulse settings The device should be filled just before use Fill Connector Appearance may vary per model Vent to Fill Lever Pulse Selector w Label PORTABLE system description Cannula Conn...

Page 8: ...WHEN LOCKING INTO POSITION PM2202 Caire Side Fill Rotate the Portable System counterclockwise until the pin of the device engages with the slot of the reservoir connector approximate rotation 45 Care...

Page 9: ...and when attempting to disconnect it Figure 2 PM2201 PM2202 Disconnect the Portable System from the Reservoir by rotating it in counterclockwise direction until the device separates The device may now...

Page 10: ...rk DO NOT USE FORCE Simply allow a few minutes for the frozen parts to warm then disengage the Portable when the ice has melted 11 Check the approximate oxygen contents in the Portable System using th...

Page 11: ...the contents scale and allow it to snap back into the scale doing so will damage the scale and void the warranty The table below shows approximate use times for the Portable System after it has been...

Page 12: ...vice to occasionally skip a breath or to pulse twice in one breath 5 Breathe through the nose and feel a pulse of oxygen as you inhale 6 Insert Portable System into the carry bag Carry bag may become...

Page 13: ...Placing the device on its side or upside down will shorten the usage time device Figure 6 Using the Carry Bag Figure 7 illustrates how the device should be placed in its bag Ensure the device is orien...

Page 14: ...ent Replacement parts Description Part Cannula 504833 Carry Bag Liner 504488 Carry Bag with Liner 504392 Contents Scale 504393 MAINTENANCE ATTENTION Equipment Provider This device contains several fie...

Page 15: ...n 1 Refill device 2 Set to prescribed setting 3 Rotate Pulse selector to your prescribed setting 4 a Check position of cannula in nose b Do not breathe through mouth 5 Reconnect cannula 6 Remove kink...

Page 16: ...Allow time for device to warm D Device frosted no pulse 1 Device overfilled 1 Wait approximately 30 minutes until unit returns to normal operating conditions E Device is making a hissing noise venting...

Page 17: ...16 NOTES device serial In Service Date Date...

Page 18: ...ions have been made to the goods shall in its discretion and at its own expense repair or replace the defective component s ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES AND THERE ARE NO WARRANTIES OTH...

Page 19: ...spiratory Therapy Equipment Part 2 Tubing And Connectors 2002 EN 1418 Welding personnel approval testing of welding operators for fusion welding and resistance weld setters for fully mechanized and au...

Page 20: ...n USA ISO 13485 Certified Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Visit us on the Web www precisionmedical com Toll Free Tel 800 272 7285 Toll Free Fax 800 353 1240 Tel 001 610 2...

Reviews: