5. Using a clean, dry cloth, wipe the fill connector on the reservoir and
Device.
6. carefully position the Portable , ensuring that the fill connector of the
Portable System aligns with the fill connector of the reservoir.
7.
engagement
:
PM2200
(Puritan bennett, Top fill)
:
•connect the Portable & reservoir by
pressing down to the fill position, being
careful not to depress the release button
on the reservoir.
•During filling, maintain a slight downward
pressure on the Portable System with
one hand to keep the device steady and
maintain proper filling position.
PM2201
(Mark Series, Top fill with
Twist):
•rotate the Portable System clockwise
until the device is locked into position
(approximate, rotation 90°) .
Do NoT USe eXceSS Force WHeN
lockINg INTo PoSITIoN.
PM2202 (Caire, Side fill):
•rotate the Portable System
counterclockwise until the pin of the
device engages with the slot of the
reservoir connector (approximate rotation
45°).
•carefully and firmly rotate the Portable
System back to the upright position,
until the Device and reservoir are locked
together.
7
WARNINg
dO NOT
fill the device while it is in the carrying bag.
The Portable System is intended to be used with any compatible
reservoir with a maximum working pressure of 50 PSI (3.45 bar).
filling the Portable System from the Reservoir
1. check the contents indicator on the reservoir to ensure liquid
oxygen is available for filling purposes. When the reservoir is low,
inform your equipment Provider.
2. remove the cannula from the Device, if attached.
3. remove the Device from the carrying bag.
4. remove protective cap on reservoir fill coupler, if applicable.
OPERATINg INSTRUCTIONS
continued