Portable Oxygen Concentrator
SPECIFICATIONS (continued)
Oxygen Minute Volume*: Setting 1:
240 +/-15% mL/min
Setting 2:
380 +/-15% mL/min
Setting 3:
520 +/-15% mL/min
PM4150 only
Setting 4:
660 +/-15% mL/min
Setting 5:
780 +/-15% mL/min
Oxygen Concentration:
87% to 95% on all settings
Mode of Delivery:
Pulsed Dose only
Trigger Sensitivity:
0.50 cmH
2
O maximum
Trigger Delay:
0.11 seconds typical
Power:
AC Power Adapter:
100-240 VAC (1.0 A @ 120 VAC)
DC Power Adapter:
11-16 VDC (5.0 A @ 12 VDC)
Battery Duration:
PM4130
PM4150
(Approximate)
Setting 1:
5.5 hrs
Setting 1:
4.7 hrs
Setting 2:
4.0 hrs
Setting 2:
3.4 hrs
Setting 3:
3.0 hrs
Setting 3:
2.7 hrs
Setting 4:
2.0 hrs
Setting 5:
1.6 hrs
Times may vary based on atmospheric conditions and battery age.
Sound
(@ Setting 2):
PM4130
- 39 dBA
PM4150
- 44dBA
Cannula:
Maximum 7 ft (2.1 m) long standard adult single lumen nasal cannula.
Equipment Classification:
Class II
- with respect to protection from electrical shock
Type BF
- degree of protection against electrical shock
IP21
- Protected from touch by fingers and objects greater
than 12 millimeters. \ Protected from condensation.
Volatile Organic Compound (VOC) and Particulate Requirements:
The oxygen delivered from the EasyPulse POC meets the following requirements for
particulate levels, VOC levels, carbon monoxide levels, carbon dioxide levels and
ozone levels.
EPA PM 2.5
Particulate Matter
ASTM D5466
Ozone Levels
21 CFR 801.415
Carbon Monoxide Levels
EPA NAAQS - Carbon Monoxide
Carbon Dioxide Levels
OSHA Permissible Exposure Limits
Standard Test Method for Determination of Volatile Organic Chemicals in
Atmospheres (Canister Sampling Methodology)
* At Sea level
Specifications are subject to change without prior notice.