Precision Medical Easy Pulse POC PM4100 Series User Manual Download Page 10

9

Portable Oxygen Concentrator

CAUTION

•  Inspect the POC for visual damage before use, DO NOT USE if damaged.

•  To

 

prevent damage to the POC, 

DO NOT

 operate the POC without 

the inlet filter or while the inlet filter is wet.

Using the EasyPulse POC

1.  Verify battery status (level) prior to each use, charge as necessary.

2.  Connect the end of a standard adult single lumen oxygen nasal cannula 

with a maximum length of 7 feet (2.1 m) to the POC’s oxygen outlet fitting.

 

To ensure proper device operation, use only 

an adult single lumen cannula maximum 7 

ft (2.1 m).

3.  Place the cannula over your ears and position 

the prongs in your nose as instructed by your 

Equipment Provider or cannula manufacturer.

OR

4.  Prior to operation, ensure the inlet and exhaust vents on the POC are clear. 

Any blockage of the vents can inhibit performance.

5.  Choose a power option: AC Power Adapter, DC Power Adapter or Internal 

Battery  (Refer  to  the  “POWER  SUPPLY”  section  of  the  user  manual  for 

details on these 3 options).

6.  Turn “ON” the POC by selecting the setting for your prescribed flow.The setting 

number will illuminated green when properly selected.

7.  Breathe normally through your nose. A measured pulse of oxygen will be 

delivered each time the POC senses an inhalation. To show the POC is 

triggering properly the illuminated setting number will turn off briefly each 

time it senses an inhalation.

8.  To turn “OFF” the POC, select the “OFF” ( ) position.

WARNING

ISO 80601-2-69 (Standard for Oxygen Concentrators) highly 

recommends that the patient cannula that delivers gas to the patient 

from the oxygen concentrator should include a Fire Stop Check Valve 

to stop the flow of gas towards the patient in the case that the cannula 

becomes ignited. The Fire Stop Check Valve should be located as 

close to the patient as reasonably practicable.

Fire Stop Check Valve - Precision Medical Part # 507706

2” Tubing Connector - Precision Medical Part # 507707

Summary of Contents for Easy Pulse POC PM4100 Series

Page 1: ...ies SAVE THESE INSTRUCTIONS Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485...

Page 2: ...SAL INSTRUCTIONS 17 LIMITED WARRANTY 18 DECLARATION OF CONFORMITY 19 RECEIVING INSPECTION Remove the Precision Medical Inc EasyPulse Portable Oxygen Concentrator POC from the packaging and inspect for...

Page 3: ...t the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage 62NA PM4130 AND PM4150 WITH RESPECT TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZ...

Page 4: ...DO NOT reach for the POC that has fallen into water Unplug immediately DO NOT attempt to repair POC Any attempt could result in electrical shock POC is designed to be used in dry conditions the POC is...

Page 5: ...NOT place the POC in a small closed space such as a closet DO NOT use extension cords or electrical adapters The POC is designed to operate with a single lumen adult cannula with a maximum length of...

Page 6: ...here the oxygen concentrator is located If unable to leave the room you must wait 10 minutes after you have turned the oxygen concentrator off before smoking Open flames during oxygen therapy are dang...

Page 7: ...equipment provider are accountable for ensuring the compatibility of the oxygen concentrator and all of the parts or accessories used to connect to the patient before use The Oxygen Concentrator its...

Page 8: ...39 dBA PM4150 44dBA Cannula Maximum 7 ft 2 1 m long standard adult single lumen nasal cannula Equipment Classification Class II with respect to protection from electrical shock Type BF degree of prote...

Page 9: ...Additional monitoring or attention may be required for patients using this POC who are unable to hear or see alarms or communicate discomfort The Setting Knob does not rotate 360 Rotating the knob le...

Page 10: ...er Adapter DC Power Adapter or Internal Battery Refer to the POWER SUPPLY section of the user manual for details on these 3 options 6 Turn ON the POC by selecting the setting for your prescribed flow...

Page 11: ...e carry bag is not intended to make the POC waterproof or to protect the device from major falls or accidents abuse When inserting the POC into the carry bag make sure of the following Before placing...

Page 12: ...he POC cool and operating properly Be sure the exhaust vents are facing away from body while being carried Iftheoutsideofthebagfeelshotorcausesanydiscomfort set the POC down and check to be sure all v...

Page 13: ...rnal power sources AC Power Adapter The AC power adapter connects the POC to a 100 240 VAC 50 60 Hz wall outlet The power adapter converts AC voltage to DC voltage in order to power the POC The AC ada...

Page 14: ...and simultaneously charge the internal battery DC Plug Auto Accessory Plug Auto Accessory Plug DC Plug connected to Power Input Auto 12 V Accessory Jack Connection To connect the DC Power Adapter to t...

Page 15: ...ternal battery will begin charging as soon as power is connected You will also be able to run the POC while the internal battery charges Lithium Ion Battery Guidelines Proper Use In order to prolong l...

Page 16: ...o breath detected for 2 minutes 1 Verify the cannula is not kinked 2 Verify the cannula is properly connected to oxygen outlet 3 Verify the cannula is properly positioned in the nose 4 If alarm persis...

Page 17: ...itive beep Pressure outside of normal operation 1 Select the OFF position 2 Select the prescribed setting 3 If alarm persists change to alternate oxygen source and contact your Equipment Provider Red...

Page 18: ...e POC until it is thoroughly dry 7 When not in use store the POC in a clean area free from grease oil and other sources of contamination Inlet Filter 1 Remove the inlet filter by pulling on the protru...

Page 19: ...n expense ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES The representatives of Precision Medical Inc or any retailers are not authorized to make oral warranties about the merchandise described in this...

Page 20: ...relevant provisions of the aforementioned legislation and the result entitles the organization to use the CE 0473 marking on those products listed above Applied Standards EN 1041 EN ISO 14971 EN 6060...

Reviews: