Balance Control (factory adjust only)

LO 
Calibration 
Control

HI 
Calibration 
Control

Decimal Point
Jumper Array

Ribbon Cable

Connector

R11

LO

R10
BAL

HI

R9

JP1

Calibration

The LO control (R11) is located below the display and the HI control 
(R9) is located to the right of the display. Apply a 4 mA signal and 
adjust the LO control to display the desired reading. Next, apply a 
signal between 16 and 20 mA and adjust the HI control to display 
the desired reading. Complete the calibration by making any 
minor adjustments to the LO and HI controls.

MOUNTING 

PD660

The PD660 can be wall mounted using the mounting holes beneath 
the cover screws. It can be panel mounted with the addition of the 
PDA6604 panel mount kit. It can also be pipe mounted by using the 
PDA6845 or the PDA6845-SS 2" pipe mounting kit. 

PD661

The PD661 has no provisions for wall mounting. Installation of 
the unit is accomplished by using the two ½" NPT conduit holes 
provided with ½" NPT fittings. 

DIMENSIONS

PD660

      

 

 

 

      

Units: Inches (mm)

Wall  
Mounting
Holes 
Beneath
Cover 
Screws

D

A

C

E

F

B

A: 3.15 (80) 

C: 2.36 (60) 

E: 2.56 (65)

B: 4.33 (110) 

D: 3.54 (90) 

F: 0.79 (20)

PD661

A

B

C

A: 4.5 (114)   

  B: 4.75 (121)   

C: 3.75 (95)

SETUP

The only tools needed for calibration are a calibrated current 
source, a flat head screwdriver, and a phillips head screwdriver. 
Please note that the meter must be disassembled in order to 
perform the setup functions.

Disassembly 

The calibration controls are located behind the display faceplate. 
To access these controls, you must first remove the enclosure 
cover and faceplate by doing the following:
PD660: Loosen the four screws on the enclosure cover and remove. 
Unscrew the two fasteners that hold the faceplate, then remove.
PD661: Turn the enclosure cover counterclockwise to remove. 
Unscrew the two fasteners that hold the faceplate, then remove.

Connections

Field connections are made to the screw terminals located on 
the Input Signal Board.  To access these screw terminals it is 
necessary to remove the Display Board from the Input Signal 
Board. First, disconnect the ribbon cable connector from the 
Display Board. Next, loosen the two screws located to the left and 
right of the LCD that hold the Display Board in place. Finally, 
remove the Display Board carefully to avoid contact with any 
rough surfaces.

Connect a 4-20 mA input signal to terminal J5 located on the Input 
Signal Board as shown below.

Note: When reassembling boards DO NOT over-tighten screws.

CAUTION - DISCONNECT FROM SUPPLY BEFORE OPENING. KEEP COVER TIGHT WHILE CIRCUITS ARE ALIVE. 
CONDUIT SEALS MUST BE INSTALLED WITHIN 18" OF THE ENCLOSURE. 
ATTENTION - OUVRIR LE CIRCUIT AVANT D’ENLEVEL LE COUVERCLE GARDER LE COUVERCLE BIEN FERME TANT QUE 
LES CIRCUITS SONT SOUS TENSION. UN SCELLEMENT DOIT ETRE INSTALLE A MOINS DE 450 mm DU BOITIER.

Decimal Point Selection

Decimal point selection is accomplished using JP1 located behind 
the faceplate to the right of the display. Leave jumper on one pin 
only for a display of 1999 (default), place the jumper over both 
bottom pins for a display of 199.9,  middle for 19.99, top for 1.999

Ribbon 

Cable

Connector

Display Board

Input Signal Board

Pollution Degree 2

Installation Category II

Screw 

Terminal

Connector J5

+

+

+

-

-

-

Loop Power 

Supply

Signal Input Connections

S

S

4-20 mA

Transmitter

J5 Screw Terminal Connector 

Wiring Diagram

LIM660-661_E               11/05

www.predig.com

PRECISION DIGITAL CORPORATION

19 Strathmore Road • Natick MA • 01760 • USA • Tel: (800) 343-1001 • Fax: (508) 655-8990 