PRECISION DIGITAL PD233 Operating Instructions Manual Download Page 5

– 5 –

While in unit of measurement mode (eg: psi),

press the ZERO CLEAR button to rezero the

gauge. This feature functions when displayed

pressure is within ±5% of zero value

This bargraph indicates 

battery level; the more 

segments, the closer the 

battery is to full charge

Flashing display when unit

pressured beyond full-scale

Press to turn backlite 

on or off

Range on keypad;

complies with ASME B40.7

Press to turn unit on or off

AUTO OFF

(Turns unit off after 

option selected)

Never*

2 minutes

5 minutes

15 minutes

30 minutes

CALIBRATE

Zero and span 

adjustments,

password

protected

DISABLE

Allows for 

“lockout” of 

CONFIG

options

CONTRAST

(Customizes

display

readout)

7 available

Selection 4 is

Default*

*Indicates Default

QUICK REFERENCE

– 12 –

Step 3: Enter user five digit password (PW). This

is the same password established to access the

CONFIG mode in the menu.
Step 4: Press the CONFIG key to select the desired

option.
Step 5: Press ENTER to finalize the selection.
Notes:

Selecting the DISAB feature does not disable the

CLEAR button on the keypad for the MAX/MIN fea-

ture. If the DISAB feature is selected, pressing the

ZERO button on the keypad will cause the display to

read DISAB for two seconds. The gauge will then

revert back to the unit of measure of the gauge. The

DISAB feature disables the zero feature of the gauge. 
Zero Disable Feature:

This feature allows disabling the Zero/Clear button

on the keypad. It also allows for a zero tolerance of

either 5% (default) or 10% of the gauge range.
Step 1:

Press the CONFIG key until the word ZERO

SP appears.

Step 2:

Press ENTER.

Step 3: 

A prompt appears to enter PW  (enter

password). The ZERO SP password is the same

password as discussed on page 10 and the head-

ing CALIBRAT:, Gauge Calibration.  Follow the

instructions on page 10 to enter a password.

Step 4:

Press the CONFIG key to select the zero

tolerance, either 5% or 10% of range, or press the

CONFIG key again and the word DISAB appears.

Press ENTER to select the new default setting.

Summary of Contents for PD233

Page 1: ...elow Pipe to which gauge is attached must be properly grounded 4 00 1 860 17 Battery Installation and Replacement The gauge comes standard with a quantity of three AAA alkaline batteries installed Use either Duracell MN2400 MX2400 or Energizer E92BP X92RP AAA alkaline non rechargeable batteries Batteries have a life of approximately 1000 hours Battery life is dependent on gauge usage backlite sett...

Page 2: ...er Nylon Protective Carrying Pouch Display Description Flashing Gauge over underpressured 0 or 100 beyond 105 of range LOW BAT Low battery replace ERROR Internal error call customer service 203 378 8281 Pressure unit conversion exceeds RES ERROR display resolution or gauge pressured beyond resolution SPECIFICATIONS WARNING AND ERROR MESSAGES 2 Congratulations on your purchase of the Precision Digi...

Page 3: ...F Dial Size 3 Case Material 300 Series SS Case Finish Electropolished Case Enclosure Rating Weatherproof IP65 Socket Material 316 SS Socket Size 1 4 NPT Male others on application Socket Location Lower connection 6 00 o clock Ranges Vac thru 7000 psi see other engineering units below for other units of measurement Operating Temperature 0 150 F Storage Temperature 40 180 F DISPLAY Type LCD Display ...

Page 4: ... 000 4 Bargraph of full scale Flashing display when unit pressured below zero Press to indicate minimum or maximum pressure gauge has measured Press again to return to pressure units While in max or min mode press to clear minimum or maximum displayed values UNITS Pressure ENGLISH PSI INHG INH2O with Temperature Options 60 F 4 C 20 C FTSW METRIC BAR MBAR KPA MPA mmHG CMH2O mmH2O KG CM2 UPDATE RATE...

Page 5: ...elect the desired option Step 5 Press ENTER to finalize the selection Notes Selecting the DISAB feature does not disable the CLEAR button on the keypad for the MAX MIN fea ture If the DISAB feature is selected pressing the ZERO button on the keypad will cause the display to read DISAB for two seconds The gauge will then revert back to the unit of measure of the gauge The DISAB feature disables the...

Page 6: ... default and means the unit can be rezeroed between 5 of span If DISAB is selected the zero button is deactivated and no display change will occur when the zero button is pushed Step 1 Press the CONFIG key until the word ZERO SP appears Step 2 Press ENTER 6 Turns the gauge on and off When pressing the ON OFF key while in the off position gauge start up display first indicates the software version ...

Page 7: ...e the gauge will turn itself off after no activity Indicates default 10 Step 1 Press the CONFIG key until the word CAL IBRAT appears Step 2 Press the ENTER key Step 3 The letters asterisks PW appear Step 4 Press the CONFIG key An Ø appears in the first position Step 5 Press the ENTER key once Step 6 Press the CONFIG key until Ø appears Ø will appear in the second position Step 7 Press ENTER Step 8...

Page 8: ...lt password that allows access to gauge calibration To access the factory default password KEYPAD FUNCTIONS Indicates default Indicates default 8 Offerings are Never 2 5 15 or 30 minutes To use the AUTO OFF option Step 1 Press the CONFIG key until the word AUTO OFF appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select the desired AUTO OFF rate Step 4 Press the ENTER key to final...

Reviews: