GAUGE INSTALLATION
Pipe Mount – 
The digital test gauge comes standard with a

1

4

NPT connec-

tion. Good piping practices recommend using teflon tape or a pipe sealant

on the gauge threads. Utilize a 

7

16

˝ wrench on the wrench flat of the gauge

to tighten the gauge to the process.
NEVER TIGHTEN GAUGE THREADS BY HOLDING THE BODY OF THE GAUGE.

DOING SO MAY DAMAGE THE GAUGE AND MAKE THE GAUGE INOPERABLE.
Panel Mount – The lower connected digital test gauge is available with an

optional flange for panel mounting. Please refer to illustration and dimen-

sions below.

!

Pipe to which gauge is attached must be properly grounded.

4.00

1.860

.17

Battery Installation and Replacement:
The gauge comes standard with a quantity of three AAA alkaline batteries

(installed). Use either Duracell MN2400, MX2400 or Energizer E92BP,

X92RP AAA alkaline, non-rechargeable batteries.
Batteries have a life of approximately 1000 hours. Battery life is dependent

on gauge usage, backlite settings and power off settings. When the  display

flashes LOW BAT, batteries should be replaced.
To replace the batteries:
1) Remove the single screw on the back of the gauge case. 
2) Hold the keypad in the palm of hand.
3) Carefully remove the three batteries from the holder and replace the bat-

teries. Use only AAA alkaline non-rechargeable batteries.

HOLE

CUTOUT

DIAMETER

3.406

(3

13

32

±

1

32

)

Precision Digital Corporation
19 Strathmore Road
Natick, MA 01760-2418
Tel: 800-343-1001
Fax: 508-655-8990

Summary of Contents for PD233

Page 1: ...elow Pipe to which gauge is attached must be properly grounded 4 00 1 860 17 Battery Installation and Replacement The gauge comes standard with a quantity of three AAA alkaline batteries installed Use either Duracell MN2400 MX2400 or Energizer E92BP X92RP AAA alkaline non rechargeable batteries Batteries have a life of approximately 1000 hours Battery life is dependent on gauge usage backlite sett...

Page 2: ...er Nylon Protective Carrying Pouch Display Description Flashing Gauge over underpressured 0 or 100 beyond 105 of range LOW BAT Low battery replace ERROR Internal error call customer service 203 378 8281 Pressure unit conversion exceeds RES ERROR display resolution or gauge pressured beyond resolution SPECIFICATIONS WARNING AND ERROR MESSAGES 2 Congratulations on your purchase of the Precision Digi...

Page 3: ...F Dial Size 3 Case Material 300 Series SS Case Finish Electropolished Case Enclosure Rating Weatherproof IP65 Socket Material 316 SS Socket Size 1 4 NPT Male others on application Socket Location Lower connection 6 00 o clock Ranges Vac thru 7000 psi see other engineering units below for other units of measurement Operating Temperature 0 150 F Storage Temperature 40 180 F DISPLAY Type LCD Display ...

Page 4: ... 000 4 Bargraph of full scale Flashing display when unit pressured below zero Press to indicate minimum or maximum pressure gauge has measured Press again to return to pressure units While in max or min mode press to clear minimum or maximum displayed values UNITS Pressure ENGLISH PSI INHG INH2O with Temperature Options 60 F 4 C 20 C FTSW METRIC BAR MBAR KPA MPA mmHG CMH2O mmH2O KG CM2 UPDATE RATE...

Page 5: ...elect the desired option Step 5 Press ENTER to finalize the selection Notes Selecting the DISAB feature does not disable the CLEAR button on the keypad for the MAX MIN fea ture If the DISAB feature is selected pressing the ZERO button on the keypad will cause the display to read DISAB for two seconds The gauge will then revert back to the unit of measure of the gauge The DISAB feature disables the...

Page 6: ... default and means the unit can be rezeroed between 5 of span If DISAB is selected the zero button is deactivated and no display change will occur when the zero button is pushed Step 1 Press the CONFIG key until the word ZERO SP appears Step 2 Press ENTER 6 Turns the gauge on and off When pressing the ON OFF key while in the off position gauge start up display first indicates the software version ...

Page 7: ...e the gauge will turn itself off after no activity Indicates default 10 Step 1 Press the CONFIG key until the word CAL IBRAT appears Step 2 Press the ENTER key Step 3 The letters asterisks PW appear Step 4 Press the CONFIG key An Ø appears in the first position Step 5 Press the ENTER key once Step 6 Press the CONFIG key until Ø appears Ø will appear in the second position Step 7 Press ENTER Step 8...

Page 8: ...lt password that allows access to gauge calibration To access the factory default password KEYPAD FUNCTIONS Indicates default Indicates default 8 Offerings are Never 2 5 15 or 30 minutes To use the AUTO OFF option Step 1 Press the CONFIG key until the word AUTO OFF appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select the desired AUTO OFF rate Step 4 Press the ENTER key to final...

Reviews: