Model PD6300 Pulse Input Rate/Totalizer
Instruction Manual
29
SETUP AND PROGRAMMING
The meter has been factory calibrated to read input
frequency in Hz (pulses/sec). The calibration equip-
ment is certified to NIST standards.
Use
the
K-Factor
menu to match the rate/totalizer with
a flowmeter’s k-factor (pulse/unit of measure).
Or
use
the
Scale
menu to scale the pulse input
(pulse/sec) without a signal source.
Or
use
Cal
menu to calibrate the rate/totalizer using a
signal source.
Overview
There are two switches, located to the right of the input connector, which
must be configured according to the input level and type. Jumper J4 lo-
cated inside the meter, behind the input signal connector, is used to se-
lect the excitation voltage (24 V*, 10 V or 5 V) which is supplied to the
P+ and P- wiring terminals.
Setup and programming is done through the front panel buttons.
After power and input signal connections have been completed and veri-
fied, apply power to the meter.
*
Default setting