ProtEX-MAX PD8-765 Explosion-Proof Process & Temperature Meter
Instruction Manual
16
Specifications
Except where noted all specifications apply to
operation at +25°C.
General
Display
1.20" (31 mm), red LED,
Four digits (-1999 to 9999),
automatic lead zero blanking.
Display
Intensity
Eight user selectable intensity levels.
Default intensity is six.
Display
Update Rate
Process/RTD: 3.7-5/second
Thermocouple: 1.8-2.5/second
LED Status
Indicators
See
for details.
Overrange
Display flashes
9999
Underrange
Display flashes
-1999
Programming
Methods
Four SafeTouch through-glass buttons
when cover is installed. Mechanical
buttons can be used with the cover
removed. Cloning with Copy feature.
Free PC-based USB MeterView
programming software, and Modbus
®
registers.
Noise Filter
Programmable from 2 to 199
(0 will disable filter)
Filter Bypass
Programmable from 0.1 to 99.9% of
calibrated span
Recalibration
All ranges are calibrated at the factory.
Recalibration is recommended at least
every 12 months.
Max/Min
Display
Max/min readings reached by the process
are stored until reset by the user or until
power to the meter is turned off.
Password
Restricts modification of programmed
settings
Non-Volatile
Memory
All programmed settings are stored in
non-volatile memory for a minimum of ten
years if power is lost.
Power
Options
85-265 VAC, 50/60 Hz; 90-265 VDC, 20 W max
or 12-36 VDC; 12-24 VAC, 6 W max
Fuse
Required external fuse: UL Recognized,
5 A max, slow blow; up to 6 meters may
share one 5 A fuse
Normal Mode
Rejection
64 dB at 50/60 Hz
Isolation
4 kV input/output-to-power line
500 V input-to-output
or output-to-P+ supply
6X5 models only:
100 V output-to-24 VDC supply
Overvoltage
Category
Installation Overvoltage Category II:
Local level with smaller transient
overvoltages than Installation Overvoltage
Category III
Environmental
T6 Class operating temperature range
Ta = -40 to 60°C
T5 Class operating temperature range
Ta = -40 to 65°C
Storage temperature range:
-40 to 85°C (-40 to 185°F)
Relative humidity:
0 to 90% non-condensing
Connections
Power, signal, relays, mA out: Removable
screw terminal blocks accept 12 to 22
AWG wire
RS-485: Removable screw terminal block
accepts 16 to 30 AWG wire
Enclosure
Explosion-proof die cast aluminum with
glass window, corrosion resistant epoxy
coating, color: blue
NEMA 4X, 7, & 9, IP68
Default conduit connections: Four ¾" NPT
threaded conduit openings and two ¾"
NPT metal conduit plugs with 12 mm hex
key fitting installed. Additional conduit
opening configurations may be available;
verify quantity and sizes on specific device
labeling during installation.
Mounting
Wall Mounting:
Four (4) mounting holes provided for
mounting meter to wall. See
additional details.
Pipe Mounting:
Optional pipe mounting kit (PDA6848)
allows for pipe mounting. Sold separately.
See
on page
for additional details.
Tightening
Torque
Screw terminal connectors:
5 lb-in (0.56 Nm)
Overall
Dimensions
6.42" x 7.97" x 8.47" (W X H X D)
(163 mm x 202 mm x 215 mm)
Weight
16.0 lbs (7.26 kg)
Warranty
3 years parts & labor. See Warranty
Information and Terms & Conditions on