With the transmitter and receiver both set to the same test
frequency, place the transmitter inside the enclosure on a marked
reference position near the center of the enclosure. Ensure the
transmitter output level is set to -30 dB.
Place the receiver outside the enclosure on a marked reference
position, at least four feet from the transmitter. Ensure that the
receiver reference position is within line of sight of the
transmitter.
Set the receiver to Shield Level mode, the receiver will flash
“INITIALIZE? PRESS CAL” on the display.
Ensure that there are no obstructions between the transmitter
and receiver and then depress the CAL button on the Receiver.
The display should indicate a Shield Level of 60 dB.
Return to the transmitter and depress the ATTN button. The
attenuation will toggle to 0 dB and the transmitter is now at +30
dBm output.
Verify that the Shield Level reads less than 3 dB on the receiver.
This may vary slightly as every environment is unique. A reading
of 0 dB is ideal.
Close the door to the enclosure and begin sniffing with the
Receiver.
The Receiver is now measuring and displaying Shield Level of the
enclosure.
Seam Sniffing in Shield Level Mode