18 3114102-UK
Readout at hardware error
Error search
Readout
Cause
CJC sensor error - check device temperature
CJ.ER
Defect internal CJC sensor or CJC temperature out
of allowed range**
Flash memory error - default configuration is
loaded
FL.ER
Error in FLASH (configuration)*
No communication
NO.CO
No communication
Input error - check input connection and reset
power
IN.ER
Error levels on measurement inputs*
Programming mode only - no output signal
PROG.
Offline configuration mode (3114 powered by
communications interface)***
Invalid configuration type or version
TY.ER
Configuration read from EEprom has invalid type
or rev. no.
Hardware error
RA.ER
RAM memory error*
Hardware error
EE.ER
EEPROM memory error*
Hardware error
NO.CA
Device not factory-calibrated
Hardware error
AD.ER
A/D converter error*
Hardware error
EF.SU
External Flash error*
Hardware error
IF.ER
Internal Flash error*
!
All error indications in the display flash once per second. The help text explains the error. If the error is a
sensor error, the display backlight flashes as well - this is acknowledged (stopped) by pushing the
3
button.
*
Error is acknowledged by entering the menu and saving or by resetting the device power.
**
Error can be disregarded by selecting input type different than TC.
*** Error indication does not flash. Error is acknowledged by connecting device power.
Error indications