4.10 AC Voltage
Warning
Electric shock hazard.
Use caution to avoid damage or personal injury. Maximum input voltage: 600V
AC RMS.
• Turn the rotary switch to the voltage position. Press
to switch to AC voltage.
• Connect the red test lead to the INPUT jack and the black test lead to the COM
jack.
• Connect the test leads in parallel to the circuit under test.
• Read the measured voltage on the display.
Note:
1. In small voltage ranges, the display may show a reading before leads are
connected.This is normal because of the high sensitivity of the meter. Once the
leads are connected to a circuit, the true voltage will be displayed.
2. indicates maximum input voltage is 600V AC (non-RMS)
3. If measured voltage is greater than 750V AC RMS, the meter will sound an
alarm.
4.11 DC Voltage
Warning
Electric shock hazard.
Use caution to avoid damage or personal injury. Maximum input voltage: 600V
DC.
• Turn the rotary switch to the voltage position. The default setting is DC voltage.
• Connect the red test lead to the INPUT jack and the black test lead to the COM
jack.
• Connect the test leads in parallel to the circuit under test.
• Read the measured voltage on the display.
Note:
1. In small voltage ranges, the display may show a reading before leads are
connected.This is normal because of the high sensitivity of the meter. Once the
leads are connected to a circuit, the true voltage will be displayed.
2. indicates maximum input voltage is 600V DC.
3. If measured voltage is greater than 1000V DC, the meter will sound an alarm.
EN
PPDCM80