Electromagnetic Compatibility (EMC) 

Precautions

The pro5R™ and pro5 AIRdaptive™ models 

are suitable for use in all establishments, 

including domestic establishments and 

those directly connected to the public 

low-voltage power supply network that 

supplies buildings used for domestic 

purposes. Both models have been tested 

against, and have passed the applicable 

requirements of relevant electromedical 

standards, including EN 60601-1-2:2007.

The pro5R™ models must emit 

electromagnetic energy in order to 

perform their intended function. Nearby 

electronic equipment may be affected by 

this emission. Similarly the pro5R™ and 

pro5 AIRdaptive™ models may themselves 

be affected by electromagnetic emissions 

from other equipment in the vicinity.

The pro5R™ models are intended for use 

in the electromagnetic environment 

specified below. The customer or user of 

the pro5R™ model should assure that they 

are used in such an environment:

–  Floors should be wood, concrete or 

ceramic tile. If the floor is covered with 

synthetic material, the relative humidity 

in the room should be at least 30%.

–   Mains power quality should be that 

of a typical domestic, commercial or 

hospital environment as appropriate.

–   The pro5R™ models should not be used 

during power mains interruptions

–   Power frequency magnetic fields 

should be at levels characteristic of a 

typical location in a typical commercial 

or hospital environment.

The pro5R™ models are intended for use 

in an electronic environment in which 

radiated radio-frequency (RF) disturbances 

are controlled. 

The customer or user of the pro5R™model 

can help prevent electromagnetic 

interference by maintaining a minimum 

distance between portable and mobile RF 

communications equipment (transmitters) 

and the pro5R™ or pro5 AIRdaptive™ 

model as recommended below, according 

to the maximum output power of the 

communications equipment:

 

The adequate separation distance 

to provide base immunity to RF 

disturbances is 1.0m.

Retain this user manual for your future 

reference.

 

 

 

 

 

08

Summary of Contents for Pro5M

Page 1: ...English Instructions for Use Pro5M Model Medical Device...

Page 2: ...o5HP My7M my7 my5 my3 Personal MOVE Pro4PlusM and proMOTION are registered trademarks and or trademarks of Peformance Health Systems and or their affiliates All other trademarks are the property of th...

Page 3: ...int dizzy nauseous short of breath or in pain stop exercising immedia tely and consult with your physician Body posture muscle stiffness and muscle tension are important contributing factors in your t...

Page 4: ...ectively make up the circulatory system Pulmonary circulation is the portion of the cardiovascular system which transports oxygen depleted blood away from the heart to the lungs and returns oxygenated...

Page 5: ...variation and the shape of the bones and cartilage that make up the joint Pathology Studies Claim Power Plate training can be a beneficial tool intervention to improve general well being fitness and d...

Page 6: ...ing on and off the machine Never reach into or under the machine or tip the machine on its side while it is in operation Use the machine only for the purposes described in this user manual and only wi...

Page 7: ...ssible injury from the machine s vibrations Power Plate recommends that the user not lean back on his her heels or lock his her joints or straighten his her legs when standing upright on the machine I...

Page 8: ...are used in such an environment Floors should be wood concrete or ceramic tile If the floor is covered with synthetic material the relative humidity in the room should be at least 30 Mains power qual...

Page 9: ...machine so that it is difficult to disconnect the main plug Warning To avoid the risk of electric shock this equipment must only be connected to a supply main with protective earth grounding This appl...

Page 10: ...Operation Power Plate pro5 face plate 10...

Page 11: ...in your training The Power Plate machine is very easy to operate as you can tell from a brief review of the primary buttons on the face plate 1st Main Controls Start Repeat 2nd Frequency 25 50Hz 1Hz i...

Page 12: ...unction Therefore its RF emissions are very low and are not likely to cause any interfe rence in nearby electronic equipment RF emission CISPR 11 Class B Class A with IEC61000 3 2 Complies with IEC610...

Page 13: ...nment Surge 1 kV line s t 1 kV differential Mains power quality should be that of IEC 61000 4 5 o line s mode a typical commercial or hospital environment Voltage dips short 5 UT 5 UT Mains power qual...

Page 14: ...eld strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of...

Page 15: ...42 EEC Date of manufacture Manufacturer Type B applied part DISPOSAL Do not dispose this product as unsorted municipal waste Collection of such waste separately for special treatment is necessary CB C...

Page 16: ...ises can be added When trying to accomplish weight loss cardiovascular or endurance improvements rest time can be reduced exercise time can be extended and the number of exercises can be increased to...

Page 17: ...cademies For more information please visit our website www powerplate com PersonalAdaptation Training on the Power Plate machine is like any other type of training start with light and short training...

Page 18: ...rtz 30 45 or 60 seconds Up to 9 minutes TimeSelections VibrationEnergyOutput Low or High DimensionsWxDxH 96 x 114 x 153 cm 38 x 45 x 60 SurfacePlateDimensionsWxD 96 x 114 cm 38 x 45 Weight 185 kilogra...

Page 19: ...tions are proper please check the circuit breaker or fuse for the electrical outlet that is supplying power to the machine in order to ensure that the outlet is receiving electrical power If the conso...

Page 20: ...y be obtained by contacting Power Plate through the Contact Us page online or by writing to Power Plate Performance Health Systems UK Ltd Business Design Centre Unit 115 52 Upper Street Islington Lond...

Reviews: