Calibration and Adjustment
Calibration and Adjustment
The
PosiTector
PC
probe is factory calibrated and performs an
automatic self-check each time it calculates a measurement. For many
powder measurement applications no further adjustment is necessary.
Simply
Zero
at the beginning of each shift, then measure.
Sometimes gage measurement calculations can be influenced by the
composition of the coating powder. A calibration adjustment improves
accuracy in these circumstances.
Adjustment, or Calibration Adjustment, is the act of aligning the
gage's predictions to match those of a known cured sample.
Begin by performing a
Verification of Accuracy
(pg 8) to determine
if there is a need for an adjustment. If the average of several
thickness results is outside the combined tolerances of the two
measuring instruments (
PosiTector
PC
and dry film thickness
instrument) then two adjustment techniques are available to
correct the variance (see
Adjustment Techniques
, pg 9).
User adjustments are stored in the current Cal setting. It may be
desirable to open a new Cal setting first (see
Cal Memory
, pg 11).
Adjustment Techniques
Adjustment Techniques
To ensure the gage is working properly, the user should
periodically perform an accuracy check by comparing predicted
thickness values with actual cured thickness values (pg 8).
If the average of a series of measurements by both instruments
(
PosiTector
PC
and dry film thickness instrument) is not within
combined tolerances, the
PosiTector
PC
can be adjusted using
one of two methods...
1 Pt Adjust
- The first and most common choice.
1.
2 Pt Adjust
- If a 1 Pt Adjust does not improve accuracy.
2.
See
Cal Settings
menu, pg 10 for more details.