10
Sitter Elite
Setting Alarm Mode
(Continued)
4. LED light blinks green to indicate selected mode is in
use (fig. 6). LED blinks yellow if the alarm mode is set to
“Mute” (fig. 6).
5. If alarm is set to “Voice only” or “Voice and Tone” mode
and custom recorded message does not play, re-record
message. If this does not work, remove batteries for at
least 1 hour to reset alarm to factory default. You can
then re-record the message.
The last option selected is the mode utilized when the
sensor is activated or magnet is removed from the face
plate, depending on which monitoring mode is selected.
6. If you are using the nurse call cable, check to make sure
it is plugged into the correct wall jack and the alarm.
Make sure the mode you select alerts at the nurse’s
station when the alarm is activated.
7. When “Voice only” or “Mute” are selected, you must have the nurse call cable plugged
into your alarm and the appropriate wall jack for your nurse call system. Ensure that the
nurse call system indicators (nursing station and wall light) are activated when the cable is
unplugged from the wall jack.
8. Replace the programming control protective cover, securing it into place.
To test the modes requiring nurse call interface, remove pressure from the sensor, unfasten the
chair belt sensor, activate the PIR sensor, or remove the magnet from the face plate. Check at
the nurse’s station to ensure that the indicator for the appropriate patient alarms on the nurse
call system and the wall light, indicating there is no pressure on the sensor, chair belt sensor is
unfastened, PIR sensor has detected activity or the magnet has been removed from the face plate.
H O L D / S U S P E N D
®
Sitter
™
MODE
LED
Fig. 6