PCT200 Series CT/PT Test System User Manual
9
1.2 Technical Specifications
Items
PCT200i
PCT200Ai
Output voltage range
0-120V
Output current range
0-5Arms (15A peak)
Output power
0.0001-500VArms (1500VA peak)
100
%
I
p
n
an
d
100
%
V
A
Ratio test range
30,000:1-45,000:5
Ratio test accuracy
﹡
1-2,000 0.05% Typ. 0.10%Gur
2,000-5,000, 0.08%Typ. 0.15%Gur
5,000-30,000, 0.10%Typ. 0.20%Gur
1-2,000 0.02% Typ. 0.05%Gur
2,000-5,000,0.03%Typ 0.10% Gur
5,000-30,000, 0.05%Typ. 0.20%Gur
Max. knee point voltage
45,000V
Burden test accuracy
±0.05%
±1mΩ Typ.
±0.1%
±1mΩ Gur.
at 0.05-200
Ω
, resolution 0.1
mΩ
Rct accuracy, resolution
±0.05%±1mΩ Typ.
±0.1%±1mΩ Gur.
at 0.05-200
Ω
, resolution 0.1
mΩ
Phase displacement
﹡
1min Typ. 3min Gur.
resolution 0.01min
1min Typ. 2min Gur.
resolution 0.01min
Main supply
20%~-15%/50-60Hz(Nominal)
90-260Vac/45-65Hz (Permissible)
Input power: max 2,000VA
Operation temperature
-10
℃
~50
℃
Relative humidity
≤90%, non
-condensing
Dimensions
470x200x245mm (WxHxL)
Weight
10kg/14kg
LCD display
8.4 inch, color display
Operating system
Windows
Note:
The corresponding temperature of each parameter is 23
℃
±5
℃
. It is suggested the preheat time is longer
than 15 minutes.
The parameters are ensured accurately within one year.
The symbol “
﹡
” indicates the knee voltage is more than 3V.
Summary of Contents for PCT200 Series
Page 13: ...PCT200 Series CT PT Test System User Manual 13 1 5 Block Diagram...
Page 23: ...PCT200 Series CT PT Test System User Manual 23...
Page 24: ...PCT200 Series CT PT Test System User Manual 24...
Page 29: ...PCT200 Series CT PT Test System User Manual 29...
Page 30: ...PCT200 Series CT PT Test System User Manual 30...
Page 36: ...PCT200 Series CT PT Test System User Manual 36...
Page 37: ...PCT200 Series CT PT Test System User Manual 37 TPY type CT parameter setting...
Page 38: ...PCT200 Series CT PT Test System User Manual 38 M type CT parameter setting...