6.2 Rear Panel 

 

10. 

Power socket:

 Please use the power cable supplied by PONOVO (The cable with 

special port must be used).   

11. 

Earth ground: 

It should be grounded with earth during testing. 

12

Current output:

 It is connected with the current input of the test object. 

13

Voltage output: 

It is connected with the voltage input of the test object. 

14. 

Contact signal output:

 Please refer to the terminal definition below. 

15.

 Control signal Input: 

Please refer to the terminal definition below. 

16.

 Voltage signal Input: 

Please refer to the terminal definition below. 

17. 

Current Signal Input: 

Please refer to the terminal definition below. 

There are both current/voltage output in the front and rear panels. Current/voltage 

output on either panel can be configured. 

Summary of Contents for PA30BI

Page 1: ...CH 2021 This manual is the publication of PONOVO POWER CO LTD Any form of copy should obtain the consent of it The manual represents the technology status at publication time The product information d...

Page 2: ...4 3 High Power Output 16 5 Function and Features of Voltage Amplifier 18 5 1 Basic Principle 18 5 2 Function 18 6 Panels 20 6 1 Front Panel 20 6 2 Rear Panel 22 6 3 Terminal Definition 23 7 How to Us...

Page 3: ...it s not necessary to depress high power button high power button light should be off also 5 Never operate live signal and load wires with voltage or current output to protect the safety of equipment...

Page 4: ...pment in the place having explosive gas or water vapor 14 The 500V dangerous voltage can be in the equipment and please don t remove the cover by yourself 15 Please contact the manufacture for any mai...

Page 5: ...horized by local country could be used Ground The Instrument The instrument must be grounded safely with the supplied ground cable in testing Observe All Terminal Ratings To avoid fire or shock hazard...

Page 6: ...rther operations Any maintenance adjustment or replacement especially to circuits or accessories must be performed by authorized personnel Keep Well Ventilation Inadequate ventilation may cause increa...

Page 7: ...ed towards the air outlet Electrostatic Prevention Operate in an electrostatic discharge protective area environment to avoid damages induced by static discharges Always ground both the internal and e...

Page 8: ...ful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the in...

Page 9: ...e available to all kinds of capacitive inductive and resistive loads Generally power amplifier consists of two types of amplifiers linear power amplifier and switching power amplifier The linear power...

Page 10: ...he total power It is suggested the total power supply by PAP02 should be not less than 5 times of the total power of all the amplifiers installed on the same panel It is suggested power on each amplif...

Page 11: ...rounding conditions No dust vibration or serious electromagnetic field interferes The equipment is not recommended to be used out of laboratory Avoid the equipment to be used online out of laboratory...

Page 12: ...ndard 0 7 07V RMS 10Vpek Gain 4 5A V Option 0 11 31V RMS 16Vpek Gain 3A V Input signal Digital optional Support Aurora protocol Current accuracy 0 2 0 5 30A Current typical accuracy 0 1 Differential i...

Page 13: ...ut Input signal Gain Standard 0 7 07V RMS 10Vpek Gain 40V V Option 0 11 31V RMS 16Vpek Gain 24V V Voltage accuracy 0 2 5V 250V Voltage typical accuracy 0 1 Differential input impedance 20k Harmonic di...

Page 14: ...PA30Bi H Power Amplifier User Manual 14 3 6 Digital Display of Voltage Current Output The RMS of current or voltage output can be displayed in real time on the panel...

Page 15: ...tion detection circuit 4 Protection circuit by which the amplifier can be terminated when its temperature is too high and will resume working when it is cooled down to a reasonable degree 5 Cooling fa...

Page 16: ...at alarm light on the front panel will be on 4 3 High Power Output With high power control button depressed after several seconds the high power control button light will turn on By then the current a...

Page 17: ...30Bi H Power Amplifier User Manual 17 4 Never apply the high power output function when DC current is produced 5 Never switch to off the high power output function when amplifier is producing any outp...

Page 18: ...ortion detection circuit When the overload or short circuit situation disappears the amplifier will resume to work automatically 4 Temperature protection circuit It will terminate the amplifier as soo...

Page 19: ...disappears the amplifier can resume to work automatically If overload or short circuit occurs the amplifier will be terminated automatically and the overload alarm light on the front panel will turn o...

Page 20: ...own based on the switch 5 Current high power Press it and light in red to output high power 6 Voltage high output Press it and light in red to output high voltage 0 250V Normally the output is 0 120V...

Page 21: ...PA30Bi H Power Amplifier User Manual 21 There are both current voltage output in the front and rear panels Current voltage output on either panel can be configured...

Page 22: ...t object 13 Voltage output It is connected with the voltage input of the test object 14 Contact signal output Please refer to the terminal definition below 15 Control signal Input Please refer to the...

Page 23: ...m signal output Voltage output overload short circuit alarm signal Pause signal output Inverse time limit signal output Control Signal Input Terminal No 1 2 3 4 5 6 Output Mode COM Function Definition...

Page 24: ...Definition VA signal input VA signal input VB signal input VB signal input VC signal input VC signal input VZ signal input VZ signal input Current Signal Input Terminal No 1 2 3 4 5 6 Channel IA IB I...

Page 25: ...t signal wires is correct before energizing the amplifier When all things are ready turn on the power of PAP02 first and then turn on the power on amplifier panel The power light will be on the run li...

Page 26: ...activate the high power function as soon as this high current test was done Note 1 Normally high power button is not activated high power switch light is off also 2 For safety never plug signal cable...

Page 27: ...unning test before delivery The test items include function test linearity load stability total harmonic distortion maximum output power amplitude and frequency characteristic input output delay step...

Page 28: ...le Amount 4 pieces The current cables to connect the amplifier output to other safety sockets of generally the current parts Colour coded voltage cables SAW0202 colour coded voltage cable Amount 5 pie...

Page 29: ...ers are shown as follow 10 1 Standard Chassis Dimension W D H 600 800 1800mm Weight 50kg Maximum load 300kg 10 2 PAP 02B Power Control Device Input 3 phase 380V 10 50Hz Protection function Open phase...

Page 30: ...P02 please carefully check if the power on each amplifier is turned off 1 Turn on the power of PAP02 first 2 Turn on the individual amplifier power in testing 3 After testing is finished please turn o...

Reviews: