A-10
A Optional Accessories
A.5.1.2 Technical Specifications
General Data
Close-Up Lenses
With the close-up unit a set of lenses for the close-up range with different
focal lengths (1, 2 and 4 diopters) is delivered. A maximum of two lenses, or
one lens and the block filter can be mounted simultaneously. The lenses for
the close-up range are also screwed directly on the video camera aperture of
the scanning head before mounting the close-up unit. The stand-off distances
of the different lenses and combinations of lenses are shown in the following
table.
Table A.1: Stand-off distances and scan ranges of the close-up lenses
Max. scanning angle in x-direction:
± 5°
Max. scanning angle in y-direction:
± 4°
Dimensions (w x h x d):
124 mm x 90 mm x 75 mm and
Weight:
0.35 kg
Close-up lens or
combination
Stand-off
distance
1
1
Measured from the front of the scanning head
Min. detail size
2
2
In the center of the video image at telephoto position and maximum stand-off distance
Max. object
size
3
3
With maximum stand-off distance
Spot diameter
Depth of focus
of the laser
4
4
Max. 3 dB signal level decay
[diopters]
[mm]
[µm]
[mm]
[µm]
[mm]
2
315 … 500
25
101 x 85
44
± 2.3
2 + 1
240 … 335
20
72 x 45
34
± 1.4
4
190 … 250
15
57 x 50
28
± 0.97
4 + 1
160 … 200
13
48 x 43
25
± 0.76
4 + 2
135 … 170
11
43 x 39
23
± 0.65
Summary of Contents for PSV-500
Page 10: ...1 6 1 Safety Information...
Page 14: ...2 4 2 Introduction...
Page 50: ...4 8 4 Making Measurements...
Page 104: ...C 2 C Declaration of Conformity...