3-12
3 First Steps
3.3
Assembly
ATTENTION!
Damage caused by unsuitable procedure!
It will be safer to handle the components of the system cabinet by two
persons.
» Get help from a second person for unpacking the system cabinet.
» Mount the instrument with a second person.
ATTENTION!
Damage of the device caused by insufficient opportunity for air
circulation!
Heat is created inside the device during operation. If the heat cannot exhaust
from the housing it leads to heat accumulation. This will overheat the
electronic parts inside the device and may result in a defect.
» Always keep free all air inlets and outlets of the device.
» Keep at least 50 mm distance to the wall if required when you set up the
device!
» If the devices are in a system cabinet: Remove all covers of the system
cabinet before switching on the devices.
» If you notice a cooling fan failure, switch off the device immediately.
ATTENTION!
Damage to the cables caused by lack of diligence!
It is important to handle the cables carefully so that the transmission quality is
not affected and the functionality of the devices is ensured.
» Do not expose the cables to high temperatures.
» Prevent the cables from being damaged.
» Do not fall beyond the bending radius of the cables. Bending radius of the
main cable (Umbilical): not below 120 mm. Bending radius of other cables:
not below 50 mm.
INFORMATION
Before starting to assemble the instrument, make sure that all system
components are in perfect condition! Check the devices for external damages
such as scratches or the like. In the case of a wrong delivery, damage or
missing parts, please inform your local Polytec representative immediately,
stating the serial number of the devices.
INFORMATION
Keep the assembly instructions for all components in a safe place.