7-2
7 Technical Specifications
Ambient Conditions
Housing
Calibration
7.2.2 Metrological Properties
Signal Output ANALOG OUT
Signal voltage output for the velocity signal
Operating temperature:
+5°C ... +40°C (41°F... 104°F)
Storage temperature:
10°C ... +65°C (14°F... 149°F)
Operating altitude:
max. 3 048 m (10 000ft)
Relative humidity:
max. 80%, non-condensing
Protection reating:
IP 64 (corresponding to DIN EN 60529)
Dimensions:
refer to
Weight:
approx. 2.6 kg
Display:
LCD, 3-line, with background lighting
Calibration recommended:
2 years
Output swing:
± 4 V
Frequency range:
0.5 Hz ... 22 kHz
Resolution of the D / A
converter:
24 bit
Output impedance:
50
Ω
Load resistance:
min. 10 k
Ω
(
−
0.5% additional error)
Overrange indicator threshold: typ. 94% of full scale value
DC offset:
max. ± 10 mV
Best resolution:
0.02
(rms)
Spurious free dynamic range
(SFDR):
> 90 dB
Harmonic distortions:
< 1% THD
Propagation delay:
typ. 1.1 ms
Calibration accuracy:
± 1% (20 Hz ... 22 kHz)