background image

3-4

3 First Steps

Rear view

The rear view of the PDV with its control elements is shown in 

FIGURE

3.2.

Figure 3.2: Rear view of the PDV

1

Analog voltage output 

ANALOG OUT

 (BNC jack)

The voltage at this output is proportional to the instantaneous vibrational velocity 

of the object. The voltage is positive when the object is moving towards the 

vibrometer optics.

2

Digital signal output 

DIGITAL OUT

 (Triax jack)

At this output the velocity information is available in S / P-DIF format (Sony /

Philips Digital Audio InterFace), e.g. to display the output signal on a PC monitor 

using a sound card.

3

Connector for 

12 V

 DC voltage (4-pin circular connector)

Socket for the connecting cable of the power supply or for the battery cable

A N A L O G

O U T

D I G I T A L

O U T

1 2   V  

m a x . 1 5   W

1

2

3

Summary of Contents for PDV-100

Page 1: ...Titel User Manual Portable Digital Vibrometer P D V 1 0 0...

Page 2: ...arranty also is invalidated in the case of the equipment being tampered with or modified without authorization To return the equipment always use the original packaging Otherwise we reserve the right...

Page 3: ...Requirements 3 2 3 3 Control Elements 3 3 3 4 Installation 3 5 3 4 1 Assembly 3 5 3 4 2 Electrical Connection 3 7 3 5 Functional Test 3 7 4 Making Measurements 4 1 4 1 Start up 4 1 4 2 Selecting Suit...

Page 4: ...s Applied 7 1 7 2 Laser Vibrometer PDV 100 7 1 7 2 1 General Data 7 1 7 2 2 Metrological Properties 7 2 7 2 3 Optics 7 5 7 2 4 Table of the Visibility Maxima 7 5 7 2 5 Dimensions 7 6 7 3 Power Supply...

Page 5: ...an action which could lead to death or serious injury if it is not avoided Intended use The instrument is intended for laboratory use and for use in an industrial environment It may only be used with...

Page 6: ...r which it was intended This means that the instrument conforms with the laser class 2 and is generally very safe It is thereby usually assumed that the eyes are protected by prevention mechanisms inc...

Page 7: ...on glasses when you have to look at the target area of the laser beam long and hard to set it up Never intentionally direct the laser beam at anyone WARNING Danger from laser light Do not use any refl...

Page 8: ...for the PDV Position The position of the laser warning labels on the PDV is shown in FIGURE 1 2 Figure 1 2 Position of the laser warning labels on the PDV Laserstrahlung Nicht in den Strahl blicken L...

Page 9: ...will invalidate the warranty None of the equipment may be operated with opened housing WARNING Danger from electrical current Intervention for maintenance and repair work may only be carried out by th...

Page 10: ...e battery or subject it to great shocks Do not use any batteries which are emitting a fluid or which show external damage 1 4 2 Safety Precautions for Loading Pay attention to the following safety pre...

Page 11: ...ing The PDV 100 is shown in FIGURE 2 1 Figure 2 1 View of the PDV In the frequency range from 0 05 Hz to22 kHz the handy PDV 100 allows you to realize measurement accuracy which you previously needed...

Page 12: ...cy modulation of the Doppler signal Figure 2 2 Signals in the vibrometer In the PDV the optics and the signal processing unit are inside the housing The demodulation of the Doppler signal is purely ba...

Page 13: ...anger from dirt Handle the front optics with care Dirt may only be removed by using a soft dry cloth an optics brush and bellows Inspection Please pay attention to the following steps when unpacking 1...

Page 14: ...he optimal operating temperature after it has been switched on The PDV therefore reaches its optimal properties after a warm up period of approx 20 minutes After that you can be sure that all componen...

Page 15: ...refer to SECTION 5 5 4 Liquid Crystal Display LCD with background lighting The display shows the settings of the vibrometer The organization of the display and how to use it to operate the vibrometer...

Page 16: ...e is positive when the object is moving towards the vibrometer optics 2 Digital signal output DIGITAL OUT Triax jack At this output the velocity information is available in S P DIF format Sony Philips...

Page 17: ...ou will find among other things the serial number of the instrument 3 4 Installation 3 4 1 Assembly The PDV has serval threaded holes on the underside to mount the laser vibrometer Two mounting holes...

Page 18: ...ed as follows 1 Assemble the tripod as described in the assembly instruction provided by the manufacturer MANFROTTO 2 Then mount the fluid stage onto the tripod as described in the assembly instructio...

Page 19: ...ue between 100V to 240V For this connect the cable of the plug in power supply with the 12V circular jack on the top of the charger and put the power supply into an earthed Before you use the PDV with...

Page 20: ...e signal level display on the display lights up the optics and the input section of the electronics are working correctly 8 Set the measurement range Velo 20mm s on the PDV 9 Connect an oscilloscope t...

Page 21: ...PDV reaches its optimal properties after a warn up period of approx 20 minutes 5 Adapt the PDV settings to your measurement task You will find detailed information on how to set the settings in SECTI...

Page 22: ...arance of the overrange indicator does not necessarily mean the measurement signal is causing overrange as noise peaks caused by brief disruptions of the optical signal level can trigger the display H...

Page 23: ...vibrations when a relatively small signal is to be measured at high frequencies Under these conditions please take care of the internal saturation of the selected measurement range refer also to SECT...

Page 24: ...t generally there are two modes of different magnitudes In this case the signal level is hardly fluctuating gray line If only one mode exists the signal level is always maximum independent from the st...

Page 25: ...the front edge of the retaining ring for the front lens refer to SECTION 7 2 5 The optimal stand off distances can be ascertained as follows Equation 4 1 i e for l 138mm the optimal stand off distanc...

Page 26: ...4 6 4 Making Measurements...

Page 27: ...ng is switched off but the readout on the display is still active 5 2 Blocking the Laser Beam The PDV is equipped with a beam shutter This can be used to block the laser beam without switching off the...

Page 28: ...5 6 1 Philosophy The PDV is operated via the display using the selection key and the key SET The main menu is shown in FIGURE 5 2 as an example of the display Figure 5 2 Exsample of the display With...

Page 29: ...ttings are described in the following Velo Here you configure the velocity measurement range in mm s full scale You will find detailed information on this in SECTION 4 2 1 LP Here you configure the cu...

Page 30: ...until the following display appears The display shows the actual temperature in degrees Celsius You exit the service mode using the selection key Generating the 2 80V test signal In service mode you...

Page 31: ...he key SET pressed and then press the selection key 2 Generate the 0V test signal by then pressing the key SET several times until the following display appears The 0V test signal is only emitted whil...

Page 32: ...approximately another 15 minutes operating time available You will find information on lithium ion battery safety in SECTION 1 4 You will find further information on the battery and on charging the b...

Page 33: ...description further procedure will be determined If the PDV has to be sent back for repair always use the original packaging and enclose an exact description of the fault 6 1 No Laser Beam If no lase...

Page 34: ...ing the focusing ring Does the signal level display react If the signal level display does not react either the optics or the input section of the electronics are faulty 2 Connect an oscilloscope to t...

Page 35: ...afety IEC EN 61010 1 2002 08 Safety requirements for electrical equipment for measurement control and laboratory use EMC IEC EN 61326 1 2006 10 EMV Safety requirements for emission and immunity electr...

Page 36: ...ting IP 64 corresponding to DIN EN 60529 Dimensions refer to SECTION 7 2 5 Weight approx 2 6kg Display LCD 3 line with background lighting Calibration recommended 2 years Output swing 4V Frequency ran...

Page 37: ...Scotchlite Tape reflective film Maximum acceleration mm s m s2 20 5 0 02 2760 100 25 0 02 13800 500 125 0 1 69000 Data format S P DIF 24bit 48kSa s Frequency range 0Hz 22kHz Propagation delay approx 1...

Page 38: ...sponse of the low pass filter High Pass Filter Filter type analog 3rd order Butterworth Cutoff frequency 3dB 100Hz 10 switchable via the display Frequency roll off 60dB dec d B 0 1 0 1 0 2 0 3 0 4 0 5...

Page 39: ...distance depends on the surface properties of the object Aperture diameter 1 e mm typ 6 5 Spot size typ m 234mm 30 1m 124 2m 250 3m 370 5m 616 each additional meter plus 130 Depth of focus3 2m distanc...

Page 40: ...r o x 1 0 0 m m f r e e s p a c e a t t h e r e a r s i d e M 6 1 4 2 0 U N C M 6 M a n u f a c t u r e d b y D 7 6 3 3 7 W a l d b r o n n G e r m a n y G m b H M o d e l N o P D V 1 0 0 x x x x S e...

Page 41: ...ns 7 3 Power Supply PDV AC Mains Connection Housing Mains voltage 100 240VAC 10 50 60Hz Output voltage 12VDC Power consumption max 20W Protection rating with protective insulation IP40 correspond to D...

Page 42: ...s 7 4 1 Mounting Plate VIB A P07 Dimensions Figure 7 3 Views of the mounting plate VIB A P07 Dimensions not specified are given in mm L O C K A D J U S T 9 1 5 8 4 0 3 4 5 0 9 8 1 0 6 0 9 5 1 2 0 2 2...

Page 43: ...ature 5 C 40 C 41 F 104 F Storage temperature 10 C 65 C 14 F 149 F Relative humidity approx 80 non condensing Protection rating IP 65 corresponding to DIN EN 60529 Dimension 110mm x 45mm x 120mm Weigh...

Page 44: ...ousing 7 4 5 Motor Vehicle Supply Cable PDV DC Connection Data Dimensions 370mm x 160mm x 150mm Weight 5 5kg incl PDV plug in power supply 1 battery and battery charger Connector 12VDC on board plug s...

Page 45: ...n FIGURE A 1 Ifrequired you can store additional PDV accessories in the pockets and under the straps of the carrier bag Figure A 1 Carrier bag PDV BS with battery battery charger and PDV 100 The elect...

Page 46: ...ck Cable to connect up the laser vibrometer CAUTION Danger from mishandling Secure the cable junction at the laser vibrometer with the knurled nut When disconnecting the cable always pull the plug and...

Page 47: ...xhaustively discharged battery the Fail LED can also initially flash Switch the battery charger off and then on again As a general rule the fault state will not occur again the second time If the faul...

Page 48: ...AC to the top of the battery charger and plug the power supply into an earthed socket 5 Press the rocker switch on the battery charger in the direction of the charge level indicators The charging pro...

Page 49: ...arged battery can be used to operate the PDV but will not attain the maximum operating time Overcharging the battery is safetly prevented through the charge electronics It is a good idea to end the ch...

Page 50: ...efore the battery is exhausted you will be given a warning The battery level indicator in the display on the PDV lights up If you continue to operate the laser it will switch itself off for safety rea...

Page 51: ...o pass back batteries after use e g at public collecting points or wherever batteries of this type are being sold NOTE Batteries must not be disposed in domestic waste A 2 Motor Vehicle Supply Cable P...

Page 52: ...A 8 A Optional Accessories...

Page 53: ...directionally sensitive counting of the passing fringe pattern The velocity component in the direction of the object beam is a function of the path difference L according to Equation B 3 For a consta...

Page 54: ...B 2 B Basics of the Measurement Procedure Interferometers of this type which are directionally sensitive are described as heterodyne...

Page 55: ...C 1 C Declaration of Conformity Appendix C Declaration of Conformity Figure C 1 Declaration of conformity for the PDV 100...

Page 56: ...C 2 C Declaration of Conformity...

Page 57: ...specifications 7 2 carrier bag PDV BS components A 1 cavity length optimal stand off distance 4 4 technical specifications 7 5 clamping screw at the focus ring 3 3 cleaning housing surfaces 3 2 the fr...

Page 58: ...erometer basics B 2 heterodyne interferometer 2 2 high pass filter analogue signal acquisition 4 3 laser vibrometer specifications 7 4 select 4 3 housing opening 3 2 PDV 100 technical specifications 7...

Page 59: ...2 mounting holes M5 at the laser vibrometer 3 5 M6 at the laser vibrometer 3 5 mounting holes inch thread at the laser vibrometer 3 5 Mounting Platte VIB A P07 dimensions 7 8 O opening the housing 3 2...

Page 60: ...cations technical laser vibrometer general data 7 1 optional accessories 7 8 power supply 7 7 specifications technical dimensions dimensions 7 6 stand off distance PDV 100 technical specifications 7 5...

Page 61: ...ls labels for the sensor head 1 4 warning on the display Batt empty 6 2 Over Temp 6 1 wave length technical specifications 7 5 weight PDV 100 technical specifications 7 2 wide range input power supply...

Page 62: ...vi Index...

Page 63: ...ierre Semard 92320 Ch tillon Tel 33 1 49 65 69 00 Fax 33 1 57 19 59 60 E mail info polytec fr Internet http www polytec fr Japan JP Polytec Japan 13th floor Arena Tower 3 1 9 Shinyokohama Kohoku ku Yo...