instrument’s end surface and by annular groove on the side surface. Switch BDOI on. Select the DER 
measurement mode with the BDG1; 

3) at set statistical  error value of  less than 10 % take  five DER  measurement readings at an 

interval of at least 30 s when there is no radiation source. Then calculate average DER value of the 
external gamma background, μSv/h, using the Formula (9.1). 

4)  place  sequentially  BDG1  onto  the  dosimetric  assembly  so  that  geometrical  center  of  the 

detector coincides with the reference point where reference DER value equals 0.8; 8.0 and 80.0 μSv/h; 

5) at set statistical error value of less than 5 % take five gamma DER measurement readings 

in every reference point at an interval of at least 10 s. Then calculate average DER value 

j

H

.

 using 

Formula (9.2). 

6) calculate relative DER measurement error Qj, % using Formula (9.3). 
7)  calculate  confidence  limits  of  permissible  intrinsic  relative  error  of  DER  measurement 

DER

, %, at confidence probability 0.95 using Formula (9.4). 

Verification  results  are  considered  satisfactory  if  values  of  confidence  limits  of  intrinsic 

relative DER measurement error for all verifiable points, calculated using the Formula (9.4), do not 
exceed the limits of permissible intrinsic relative error calculated using the Formula 

)%,

20

(

.

H

K

add

   

 

 

 

 

(9.8) 

where 

.

H

 − DER value, μSv/h; 

К – coefficient, equal 2,0 μSv/h. 

 

9.8.3.4  Check  ability  to  accumulate  scintillation  spectra  of  gamma  radiation  and  relative 

energy resolution during BDG1 operation in the scintillation spectra accumulation mode as follows: 

1) connect BDG1 to BDOI; 
2)  switch  BDOI  on  and  set  into  the  spectrum  accumulation  mode  according  to  the  chapter 

11.1.5 of the Operation Manual; 

3)  place  gamma  sources 

137

Cs  and 

57

Со  (10

4

  -  10

5

)

 

Bq  from  the  reference  spectrometric 

gamma sources set of OSGI-3 type at such a distance from the scintillation detector surface against 
the geometrical center so that the count rate was no more than 10000 cps; 

4) press START button and carry out spectrum accumulation till a well-defined spectrum will 

be  displayed  on  LCD,  or  continue  accumulation  during  100  s  at  least.  The  LCD  will  display 
accumulated spectrum; 

5) stop spectrum accumulation  by pressing the MENU button, record accumulated spectrum 

into BDOI memory and send accumulated spectrum to PC as specified in Operation Manual, chapter 
11.11; 

6)  see  information  on  working  with  spectra  saved  in  the  PC  in  “Read  Me”  file  of  the  user 

program “РМ1403 spectra”; 

7) calculate the relative energy resolution 

1

, %, using the Formula (9.5). 

Verification results are considered satisfactory if relative energy resolution 

1

, %, calculated 

using the Formula (9.5) doesn’t exceed 8.5 %. 

 
9.8.3.5 Calculate permissible intrinsic relative error of photon radiation DER measurement for 

BDG2 in the following way: 

1) connect BDG2 detection unit to BDOI; 

2) place BDG2 on the verification assembly with 

137

Cs gamma source so that the instruments’ 

panel  faces  the  radiation  source,  and  the  line  going  through  the  geometrical  center  of  the  detector 
aligns  with  the  irradiation  flow  axis.  Geometrical  center  of  the  detector  is  marked  by  “+”  sign  on 
instrument’s end surface and by annular groove on the side surface. Switch BDOI on. Select the DER 
measurement mode with the BDG2; 

3) at set statistical  error value of  less than 10 % take  five DER  measurement readings at an 

interval of at least 30 s when there is no radiation source. Then calculate average DER value of the 

Summary of Contents for BDG1-PM1403

Page 1: ...GAMMA RADIATION DETECTION UNIT BDG1 1403 OPERATION MANUAL...

Page 2: ...DG1 14 6 1 Switch BDG1 On Off 14 6 2 Operability control BDG1 16 6 3 Connect BDG1 to PC 16 6 4 BDG1 operation 17 6 4 1Photon radiation DER measurement mode 17 6 4 2Scintillation spectrum registration...

Page 3: ...examination as well as other necessary information for correct instrument operation and full utilization of its capabilities BDG1 is documented as follows according to state standard of manufacturer c...

Page 4: ...de isotope compositions measure DER of gamma radiation search of ionization sources so that to prevent unauthorized transfer of radioactive sources and materials Instrument is as well used by speciali...

Page 5: ...bar 1 1 Clamp 1 1 Handle 1 1 Cable 2 1 5 m 1 1 Cable 2 2 0 25 m 1 1 Cable 3 1 Certificate 1 Data storage Software Operation Manual2 1 1 Packaging 3 1 1 Supplied by user request by separate order 2 Ve...

Page 6: ...ceeded discontinuous signal 3 5 Threshold level setting discreteness unit of least significant indicated bit Setting range of DER thresholds from 0 01 to 100 Sv h 3 6 Energies range of registered phot...

Page 7: ...han 0 25 Sv h detects standard radioactive sources SS and alternative gamma sources according to Table 3 1 1 with probability more than 0 5 Table 3 1 1 Source type Parameter name 133 Ba 137 Cs 60 Co G...

Page 8: ...BDG1 is resistant to radiofrequency electromagnetic fields test level 4 30 V m within frequencies range 80 1000 Hz from 800 to 960 Hz and from 1 4 to 2 5 GHz when there is electromagnetic emission of...

Page 9: ...ze no more than 290x70 3 35 Reliability values mean time between failures no less than 20000 hours mean life time no less than 10 years mean recovery time no more than 60 min Manufacturer guarantees i...

Page 10: ...eral BDG1 construction Front panel bears 1 LED yellow BDG1 testing mode indication green BDG1 calibration mode indication 2 Sound audio alarm for sound alarms when set thresholds levels are exceeded 3...

Page 11: ...the BDG1 geometrical effective center 2 4 2 Handle Telescopic extension bar BDG1 is equipped with a handle Figure 4 1 to enable convenient operation and a telescopic extension bar to enable access to...

Page 12: ...cessing of measurement results rapid adaptation to radiation intensity changes measurement time is inversely proportional to the radiation intensity Data communication channel enables information exch...

Page 13: ...d mechanical injuries corrosive substances organic solvents open fire sources 5 2 BDG1 power BDG1 is powered from BDOI or PC USB POWER LED on the BDG1 front panel starts blinking Figure 4 1 position 4...

Page 14: ...ront panel of BDG1 Figure 4 1 position 1 1 emits sound audio alarm when set DER DE thresholds are exceeded 2 ALARM LED for alarming 3 BAT LED for On Off state indication or BDOI charging 4 GPS receive...

Page 15: ...of the MENU 4 MODE BACK OK open the list of BDOI operation modes Exit or return to the previous operation mode or option Directional pad keys 5 UP 6 DOWN vertical cursor scrolling along menu options...

Page 16: ...1 switches off when the cable is disconnected from it or after BDOI is switched off To switch BDOI off press key ON OFF and wait till the BDOI switches off 6 2 Operability control BDG1 Switch on BDG1...

Page 17: ...beginning Filling of the scale means the end of the calibration At the end of the calibration BDG1 enters the search mode of photon radiation sources 1 2 Figure 6 4 6 4 1 Photon radiation DER measurem...

Page 18: ...rement mode menu enables following functions Figure 6 5 Clear Statistic start of the count rate and DER measurement Save To History save the history to the BDOI memory Sound On Off audible alarm on of...

Page 19: ...e relative value of the measured count rates Figure 6 7 1 2 3 Figure 6 7 Pink scale Figure 6 7 1 means that average count rate of gamma channel is low Low Count Rate less than 200 cps Low gamma channe...

Page 20: ...ar materials medical industrial natural View Library view the radionuclides library Set Absorber To obtain more reliable results of the radionuclide composition identification it is necessary to enter...

Page 21: ...display the identification result in the upper LCD area If gamma channel load is less or more than stated one then the instrument will take more time to identify During spectrum accumulation the lower...

Page 22: ...ously measured DER values in Sv h mSv h Sv h or R h mR h R h Figure 6 11 1 1 2 Figure 6 11 Count rate current value is compared with the threshold calculated in the calibration mode If the count rate...

Page 23: ...To localize RS place BDG1 near the tested object at a distance of no more than 10 cm Move the instrument alongside the object no faster than 10 m s The closer RS is the faster the signals repetition r...

Page 24: ...Figure 10 12 5 FWHM First point Width at half height of first calibration point in keV Recommended initial value is 38 0 for 122 keV energy Energy First point First calibration point energy Recommend...

Page 25: ...25 1 2 3 4 5 6 Figure 6 12...

Page 26: ...ty and reliable operation during its long service life Preventive maintenance includes external examination dust removal and deactivation if radioactive dust gets on BDG1 To deactivate BDG1 wipe it wi...

Page 27: ...DE LED doesn t light when BDG1 is connected to BDOI or PC Cable is faulty BDOI or PC is switched off Replace the cable Switch BDOI or PC on 2 Sound alarm doesn t work Sound is OFF Detector is faulty S...

Page 28: ...ton radiation DER permissible intrinsic relative error 9 8 3 1 9 8 3 3 9 8 3 5 Yes Yes Yes Yes Yes Yes estimation of the neutron radiation DER permissible intrinsic relative error 9 8 3 7 Yes Yes esti...

Page 29: ...reference sources calibration no more than 4 9 8 3 2 9 8 3 4 9 8 3 2 9 8 3 4 Thermometer Scale interval 1 0 Measurement range 10 40 9 6 1 9 6 1 Barometer Scale interval 1 0 kPa Measurement range 60 12...

Page 30: ...rnal examination must prove that instrument meets following requirements delivery kit corresponds to Operation Manual requirements there is a note in the Operation Manual about primary verification or...

Page 31: ...l center of the detector is marked by sign on instrument s surface Switch BDOI on 4 at set statistical error value of less than 10 take five DER measurement readings at an interval of at least 60 s wh...

Page 32: ...e MENU button and record accumulated spectrum into BDOI memory under the selected number as specified in Operation Manual chapter 11 11 5 send accumulated spectrum to PC according to the chapter 11 11...

Page 33: ...ation mode according to the chapter 11 1 5 of the Operation Manual 3 place gamma sources 137 Cs and 57 104 105 Bq from the reference spectrometric gamma sources set of OSGI 3 type at such a distance f...

Page 34: ...switch BDOI on switch DE measurement mode on and set maximal DE thresholds value 2 place BDG2 on the verification assembly so that geometrical center of the detector coincides with the reference point...

Page 35: ...ent error Qj using the Formula oj oj ijB j H H H Q 9 13 6 calculate permissible limits of intrinsic relative measurement error at confidence probability 0 95 using the Formula 2 max 2 max 2 3 j j o s...

Page 36: ...insic relative error add calculated using the Formula 20 A add 9 19 where coefficient equal to 10 min 1 cm 2 measured value of radiation flux 9 8 3 9 Calculate intrinsic relative measurement error of...

Page 37: ...at confidence probability 0 95 using Formula 9 18 Verification results are considered positive if values of confidence limits of intrinsic relative measurement error of radiation flux density of BDAB...

Page 38: ...ard dependence of the upper limit of the measuring range of DER of gamma radiation energy E scintillation detection channel 0 20 40 60 80 100 120 0 500 1000 1500 2000 2500 3000 3500 Energy gamma radia...

Reviews: