Description of the function 

 

 Description of the function 

 
 

The energy your lifter needs for raising and lowering the patient 
is provided by a rechargeable battery. The yellow LED at the head 
end of the lifter enables you to monitor the state of charge of 
the battery. When the manual switch is operated for the upward and 
downward movement, the LED can light briefly; this is perfectly 
normal. If the LED lights permanently, the battery must be re-
charged as soon as possible.  
To do this, grasp the battery box by the handle, remove it from 
the bracket on the lifter by tilting it slightly to the rear, and 
insert it in the bracket of the charger. 
Make sure that the battery box engages properly in the socket and 
lay the safety strap across the battery. As soon as the mains 
power supply is connected, the green LED at the charger lights to 
indicate that power is being supplied to the charger. The red LED 
lights as long as charging current is flowing. If the charging 
current drops below 40mA during the charging process, the red LED 
is extinguished to indicate that the battery is fully charged. The 
charging period lasts at least 12 hours. The battery can be re-
charged at any time. 
After charging, remove the plug from mains power supply. 
The red emergency off mushroom button serves down the lifter 
drive. 
Pressing the mushroom button brings the switch into the off posi-
tion and fixes it there mechanically. Turning the button clockwise 
releases the locking mechanism and the switch turns the drive back 
on again. In order to avoid a panic situation with the patient in 
the raised position, the small black button can be pressed in or-
der to lower the lifter. This function is only available while the 
emergency off button is in the operating position. 
 
 

Note: 

 

The motor and gearing are lubricated for life and re-
quire no maintenance. The lifter column is lightly 
greased to allow friction-free operation. 

 

 
 
 

The electrical lifter controller switches off if: 
 

1.  There is an overload of more than 1500N. 
 
2. 

 

The battery state of charge is very low (yellow LED 
lights). 

 
3.  Resistance is felt while the lifter is being lowered. 

 
 

Summary of Contents for GK-Hektor

Page 1: ......

Page 2: ...1 ...

Page 3: ...ntents Contents Contents 2 Components 3 4 User manual 5 7 Transfer from bed 5 Transfer to bed 5 Bathing procedure 6 Cleaning 6 Description of the function 7 Safety instructions 8 Service 9 Technical data 10 ...

Page 4: ...ponents 3 1 3 8 7 6 4 2 5 Components 1 Pedicure equipment 2 Safety belt 3 Backrest small 4 Push bar 5 Console and battery pack 6 Scale 7 Steerable roller with brake 8 Steerable roller with tracking device ...

Page 5: ...Components 4 Emergency off button Emergency lower ing button Hand controller with Up Down function Battery pack ...

Page 6: ...the seat 6 Close the swivel bow 7 Lift the seat 8 Release the locking device 9 Pull lifter and patient away from the bed and transfer to the bathroom Note It is recommended that the patient holds onto the sur rounding rail and the safety strap is fitted during pa tient transport Transfer to bed 1 Lower the lifter onto the upper edge of the mattress 2 Apply the locking device to the wheels 3 Open t...

Page 7: ...he swivel bow 11 Lower the bath tub completely 12 Release the locking device from the wheels and pull the lifter away from the tub Note When raising the lifter and the bath care must be taken that the patient keeps its arms folded danger of being pinched Cleaning The lifter is cleaned above the bath tub 1 Position the lifter above the bath tub 2 Remove the mattress from the stretcher if available ...

Page 8: ...he charging current drops below 40mA during the charging process the red LED is extinguished to indicate that the battery is fully charged The charging period lasts at least 12 hours The battery can be re charged at any time After charging remove the plug from mains power supply The red emergency off mushroom button serves down the lifter drive Pressing the mushroom button brings the switch into t...

Page 9: ...ge at regular intervals regardless of how often it is used We also recommend that you ask our service Department to carry out a safety and functional check at least once every year Note The safety rules and instructions in this operating ma nual must be observed in order to ensure safe and proper use of the equipment Adjustments and repairs must be carried out by authorized and appropriately quali...

Page 10: ...cted to plug Connect to plug Lifter cannot be moved downwards without applying force Pinching protection has been triggered Gently push the Lifter downwards No movement is possible although the battery is fully charged Defect fuse Unscrew the cap and check the fuse on the electronic Also check the fuse in the battery box Note These tasks must be performed only by trained and authorized personnel m...

Page 11: ...electric shock Typ B Transport and storage 20 C at 40 C Operating temperature 10 C at 40 C Dimensions 116 x 68 x 95 HxBxT cm EMC tested in accordance with EN 60 601 1 2 Charger Type DL 12 1 Power supply voltage 230V 50Hz Charging power 13 6V 1A Protection class II Loading characteristic IU Protection against ingress of moisture IP 20 Tested in accordance with EN 60 335 2 29 Disposal of the lifter ...