Polaris GK-Hektor User Manual Download Page 11

Technical data 

10 

 

 Technical data 

 
 

For the Patient lifter — seat type “Hektor”  
 
 

Lifting 

Power 

     1500N 

- Own weight 

     

 

 

 

650N 

Lifting 

height     630mm 

- Lifting speed  

  

 

 

 

20mm/sec. 

- Electrical power   

 

 

 

lead accumulator, gas-tight 

  

12V, 

7.2Ah 

 

 

Battery fuse 20Atr 

 

 

Overload protection at 14A 

 

 

Loading time 12 hours 

- „On“ period 

  

 

 

 

 

1 Min. / Pause 10 Min. 

- Tested in accordance with  

 

EN 60601-1 und Nr. 93/42 EWG 

- Degree of protection   

 

 

IP 24 

- Protection against  

electric 

shock     Typ 

- Transport and storage   

 

 

- 20°C at + 40°C 

- Operating temperature  

 

 

+ 10°C at + 40°C 

- Dimensions 

 

 

 

 

 

116 x 68 x 95 (HxBxT) cm 

- EMC tested in accordance with  

EN 60 601-1-2 

 

Charger 

 

      Type 

DL 

12/1 

- Power supply voltage   

 

 

230V, 50Hz 

- Charging power  

 

 

 

 

13.6V, 1A 

- Protection class   

 

 

 

II 

- Loading characteristic  

 

 

IU 

- Protection against ingress 

of 

moisture 

 

     IP 

20 

- Tested in accordance with  

 

EN 60 335 / 2-29 

 
 
 

Disposal of the lifter at the end of its lifetime: 
 
 

Metals 

 

     Iron 

galvanized 

coated, 

   

   

 

 

 

 

 

aluminium anodized, copper                     

 

      with 

PVC, 

high-grade 

steel 

Plastic 

     Epoxy 

with 

glass-fibre, 

 

 

      polyurethane 

Battery 

     Lead 

with 

gel-type 

electrolyte 

 

 

      (sulphuric 

acid)     

   

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for GK-Hektor

Page 1: ......

Page 2: ...1 ...

Page 3: ...ntents Contents Contents 2 Components 3 4 User manual 5 7 Transfer from bed 5 Transfer to bed 5 Bathing procedure 6 Cleaning 6 Description of the function 7 Safety instructions 8 Service 9 Technical data 10 ...

Page 4: ...ponents 3 1 3 8 7 6 4 2 5 Components 1 Pedicure equipment 2 Safety belt 3 Backrest small 4 Push bar 5 Console and battery pack 6 Scale 7 Steerable roller with brake 8 Steerable roller with tracking device ...

Page 5: ...Components 4 Emergency off button Emergency lower ing button Hand controller with Up Down function Battery pack ...

Page 6: ...the seat 6 Close the swivel bow 7 Lift the seat 8 Release the locking device 9 Pull lifter and patient away from the bed and transfer to the bathroom Note It is recommended that the patient holds onto the sur rounding rail and the safety strap is fitted during pa tient transport Transfer to bed 1 Lower the lifter onto the upper edge of the mattress 2 Apply the locking device to the wheels 3 Open t...

Page 7: ...he swivel bow 11 Lower the bath tub completely 12 Release the locking device from the wheels and pull the lifter away from the tub Note When raising the lifter and the bath care must be taken that the patient keeps its arms folded danger of being pinched Cleaning The lifter is cleaned above the bath tub 1 Position the lifter above the bath tub 2 Remove the mattress from the stretcher if available ...

Page 8: ...he charging current drops below 40mA during the charging process the red LED is extinguished to indicate that the battery is fully charged The charging period lasts at least 12 hours The battery can be re charged at any time After charging remove the plug from mains power supply The red emergency off mushroom button serves down the lifter drive Pressing the mushroom button brings the switch into t...

Page 9: ...ge at regular intervals regardless of how often it is used We also recommend that you ask our service Department to carry out a safety and functional check at least once every year Note The safety rules and instructions in this operating ma nual must be observed in order to ensure safe and proper use of the equipment Adjustments and repairs must be carried out by authorized and appropriately quali...

Page 10: ...cted to plug Connect to plug Lifter cannot be moved downwards without applying force Pinching protection has been triggered Gently push the Lifter downwards No movement is possible although the battery is fully charged Defect fuse Unscrew the cap and check the fuse on the electronic Also check the fuse in the battery box Note These tasks must be performed only by trained and authorized personnel m...

Page 11: ...electric shock Typ B Transport and storage 20 C at 40 C Operating temperature 10 C at 40 C Dimensions 116 x 68 x 95 HxBxT cm EMC tested in accordance with EN 60 601 1 2 Charger Type DL 12 1 Power supply voltage 230V 50Hz Charging power 13 6V 1A Protection class II Loading characteristic IU Protection against ingress of moisture IP 20 Tested in accordance with EN 60 335 2 29 Disposal of the lifter ...