8
4.11
Sensor operating conditions: normal ambient temperature
4.12
Pressure: 700-1060hPa
4.13
Transmitter, sensor, transmitter charger and power adapter are all suitable for patient use. The patient can be
the intended operator.
4.14
CT3 Clinical data
Main performance data
Glucose Range
Percent
within
20/20% EKF
(%)
A+B area
ratio of
Clarke error
grid
A+B area ratio of
Consensus error grid
MARD%
Full Glucose Range
91.49%
99.74%
99.95%
9.07%
5.
Operating Procedures
5.1
Prepare your transmitter
5.1.1 Clean your transmitter (if needed): Clean the electrode contact pins (Figure 2) and the surface of the
transmitter.
Figure 2 Transmitter Contact Pins
5.1.2
C
harge your transmitter: Place the transmitter into the charger to start charging; A bule
color indicates charging completion. Make sure that your transmitter is fully charged before every use.
5.2 Prepare your mobile phone
For first time users, scan the QR code on the product packaging,
search the App Store on your phone,
locate "AnytimeWell" and download it.
5.2.1 Log in
Install AnytimeWell and open the App. If it is the first time, follow the on-screen prompt to create your account.
Use your email address as your account number and create a password to complete registration (Figure3). For
normal use, log in your account by entering your email address and password. Then click Login to log into your
account.