9.5
Temperature not stable
- Restart system (option from menu, see menu item
"Restart" page 19 )
- Place top lid on the bath.
- Check PID settings. When PID settings are incorrect
the system can start to be unstable. Graph (SP-40
PID25-200-0) shows unstable control at -40°C due to a
high setting of I=200. Lowering this value to 20 results
into control of better than ±0.02°C
Standard PID setting:
Pb*=25
I=16
D=0
*Pb= proportional band. Proportional value P is found to
be 100/Pb)
-
System connected to RS232. PID settings “4” are
used. Change settings with Tamcom to Pb=25, I=16,
D=0.
- Check mains supply for stable power. Graph below
shows heavy disturbance due to fluctuations in mains
supply. Temperature accuracy is heavily disturbed by
voltage fluctuations due to switching on and off heavy
apparatus at mains supply. Previous graph(SP-40
PID=25-200-0) and graph below (SP-40 PID=25-20-0)
are same apparatus.