7.1.3
Menu item "Max Power"
(press: low, med, hi, max)
Limits the applied power by a maximum value:
Low
Maximum of 25% applied
Medium
Maximum of 50% applied
High
Maximum of 75% applied
Maximum
100% power is applied
Limiting the power can be used to lower overshoot
temperature..
7.1.4
Menu item "Boost heater"
A secondary heater is used to quickly heat up the bath.
This menu item enables or disables the boost heater.
Standard value: On
7.1.5
Menu item "Time const"
Used to select time to tune. The option precise has to be
used to reach maximum temperature accuracy. Options
are:
- Fast
60 seconds
- Medium
120 seconds
- Slow
180 seconds
- Precise
240 seconds
Standard value:
Precise
7.1.6
Menu item "Stirrer" (optional)
Optional: Stirrer 0 .. 100% (step 6%). Inactive
7.1.7
Menu item "Low alarm" (optional)
Optional: min SP to max SP. resolution 0.1°C.
No hardware connected, display function only.
7.1.8
Menu item "High alarm" (optional)
Optional: min SP to max SP. resolution 0.1°C.
No hardware connected, display function only.
7.1.9
Menu item "PID parameter"
PID set 1
-
PID set 2
-
deactivated
PID set 3
-
deactivated
PID set 4
-
Activated when communication
via RS232
Each set offers individual PID settings for:
Proportional band value
Integrating value
Differentiating value
Default settings
Pb*
:=
25
I
:=
16
D
:=
0
*Pb= proportional band. Proportional value P is found to be 100/Pb.