www.tamson.com
Tamson Instruments bv
Page 17/19
ISO 9001 : 2015
NL/PRO 238239125
van 't Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tcc-b-man Rev 1.01.docx Rev 1.01 UK
2019
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
13 Trouble shooting
:
Electronics seem to be "dead"
:
Check fuse
☺
:
Replace fuse. Fuse can be accessed
by removing right side panel.
:
Compressor (C1 or C2) do not run
:
Check apparatus for dust.
☺
:
Remove all dust using a vacuum
cleaner.
Do not use pressurised
air.
:
Fan makes noise
:
Check fans inside for blocking
☺
:
Remove dust or litter sucked in.
:
System does not cool down
:
Ice forming (crystals) in the bath.
☺
:
Use new methanol
Do not use ethanol
FUSE