background image

23                                                                                              5.7 Significant Change 
 
power in power factor studies. A revenue meter that doesn’t total negative 
power, or doesn’t include the effects of harmonics, may show readings that 
differ from this report. 

 
5.7 Significant Change 

 
The Significant Change record type tracks quick fluctuations in the line volt- 
age, with single-cycle response, while ignoring gradual changes. Voltage events 
are time-stamped to the second, and listed in a report. If the report is empty, 
there were no voltage events that exceeded the trigger threshold. This is a 
quick way to gauge the voltage power quality, because only voltage fluctua- 
tions exceeding the threshold are listed. 
 

5.7.1 Trigger Logic 

 
The Significant Change record type uses a voltage threshold parameter. At the 
end of each second during the recording session, the largest and smallest RMS 
voltages for that second are compared with the “standard” Significant Change 
voltage. This standard voltage starts as the nominal voltage picked by the 
Scanner during the two minute countdown (typically 120, 208, 240, 277, or 480 
volts). If the difference between the standard voltage and either the maximum 
or minimum voltage was more than the threshold, a Significant Change is 
recorded. In addition, the voltage (either the max or min) that caused the 
trigger becomes the new “standard” until the next Significant Change. 
As an example, consider a “standard” voltage of 119 volts, and a threshold 
of 2 volts. After 40 seconds, the voltage drops to 118 volts. No Significant 
Change is recorded because the 1 volt change is smaller than the 2 volt thresh- 
old. After another 35 seconds the voltage increase to 120 volts. The change is 
2 volts, from 118 to 120, but no Significant Change occurs because 120 volts is 
only 1 volt greater than the “standard” of 119. After another 23 seconds the 
voltage increases to 121 volts. A Significant Change is triggered because the 
1 volt increase created a 2 volt difference between the 121 maximum voltage 
for that second, and the 119 volt standard. The standard voltage is now set 
to 121 volts, until the next Significant Change. 
Only one Significant Change per second can be recorded per channel. If 
both the single-cycle max and min meet the threshold in the same second, the 
voltage that is furthest from the standard become the new standard. 
 
 
 
 
 
 
 

Summary of Contents for 2SX Series

Page 1: ...oltage range is 0 to 290 volts allowing the measurement and recording of severe voltage variations on 120VAC single phase lines Inside its lightweight rugged enclosure state of the art electronics mea...

Page 2: ...ly remove the revenue meter before attempting to remove the Scanner from the meter socket The RJ 11 serial port is optically isolated from the 120VAC line However to insure user safety and prevent dam...

Page 3: ...data files and documentation for Winscan and all Scanners in Adobe Acrobat format Acrobat reader and the latest firmware for all Voltage Scanners 3 2 Installing the PC software Winscan Insert the CD R...

Page 4: ...tallation site using normal safety pre cautions Make whatever notations your procedures require about the status and identification of the meter at the time of removal Attach the ground clip exiting t...

Page 5: ...Scanner when it is plugged in The 2SX Series Scanner can detect that it has been unplugged from a meter base and will not continue to record under battery power if it is removed from service as oppose...

Page 6: ...nicate with Winscan running on a PC or laptop to receive its initialization and to download data Use the supplied S Series Scanner serial cable to connect the 2SX Series Scanner to a PC Plug the DB9 c...

Page 7: ...indows 98 the infrared port on laptops is not compatible with the infrared port on the 2SX Series Scanner A serial cable must be used to communicate directly with a laptop 4 7 Using the LDU The LDU ha...

Page 8: ...will show you the type of hookup wye or delta the recording interval and the threshold settings Press the BLUE key to go back Move to the DIA option using the arrow keys and press SET to select it The...

Page 9: ...n menu Once the scanner is powered and the PDA is communicating with the scanner tap the PalmScan icon to reach the Main Menu consisting of 3 commands LDU Comm and Files 4 8 1 LDU This mode allows the...

Page 10: ...K to initiate the transfer Depending on the size of the file the process could take several minutes The download is complete when the Main Menu appears Back at the office the PDA may be connected to t...

Page 11: ...s in the meter adapter housing 4 9 2 Power The power option allows the 2SX Series Scanner to compute and record the power consumed by the loads on the service Real power watts reactive power vars appa...

Page 12: ...uffix giving an iVS 2SP Adding Harmonics creates an iVS 2SX PH Memory adds an M and the Battery option adds a B to the model Thus a fully loaded 2SX Series Scanner would be an iVS 2SX PHMB Since the p...

Page 13: ...re important distur bances Although it is often possible to use regulatory limits or other known standards to set thresholds this can be a chicken and the egg type problem sometimes you need to know s...

Page 14: ...aph you can see power quality events such as single cycle voltage sags and current surges as well as long term voltage trends With the graph an entire recording session can be examined at a glance 5 2...

Page 15: ...power and power factor With the harmonics option the 2SX Series Scanner can record stripcharts of harmonic magnitudes Typically only a few stripcharts are needed at one time All the Stripcharts share...

Page 16: ...left to record for days since it would only hold the last few hours of Stripchart data The best use in this case is to set the Stripchart to one second then cycle the load such as the elevator for a w...

Page 17: ...seful to see if harmonics are present The total recording time is shown by Winscan as Stripcharts are enabled and disabled during the Scanner setup Another method to increase Stripchart memory is to r...

Page 18: ...to reflect average line conditions the more loads included in the recording the better the average Monitoring a single small load such as a small office building would not create a very good profile o...

Page 19: ...of the time Where the Daily Profiles give average current power factor etc for distribution profiling the histograms show what values are the most common the mode in statistical terms 5 4 1 What s Rec...

Page 20: ...ng at the nearby bins you know how many cycles the voltage was near those extremes For example if all the bins below 110 volts are zero then you immediately know that there was not even a single cycle...

Page 21: ...minute then the Stripchart voltage averages will match the Minute Histogram counts Like the Cycle Histograms there are no settings for the Minute Histogram All available Minute Histograms in a Scanner...

Page 22: ...n compute power 5 6 1 What s Recorded Each cycle the real reactive and apparent power values are computed and added to the running totals for the recording session These values include the effects of...

Page 23: ...or 480 volts If the difference between the standard voltage and either the maximum or minimum voltage was more than the threshold a Significant Change is recorded In addition the voltage either the ma...

Page 24: ...cant Change if the sag is greater than the threshold If this happens the sag voltage becomes the standard which will trigger another Significant Change if the voltage returns its previous level The Si...

Page 25: ...nominal voltage selected by the Abnormal Voltage record type during the two minute count down The Event Change Nominal voltage is specified by the user and is not picked by the Scanner The second para...

Page 26: ...ettings are separate for each channel Another channel may trigger its own Event Change while other channels have running Events resulting in overlapping Events 5 8 2 What s Recorded When an Event Chan...

Page 27: ...report Usually Event Change is not the first report to analyze in a Scanner recording due to its complexity Check the voltage Stripchart for min or max voltages out of tolerance or the Significant Cha...

Page 28: ...fferent for various Scanners but every Scanner with Event Change can record hundreds and most over one thousand records 5 9 Power Outage The Power Outage report lists the date and time of all outages...

Page 29: ...arge enough Flicker is computed once per second based on the previous second s one cycle max min and one second average RMS voltage levels The thresholds are given as a percentage If the max min or av...

Page 30: ...the low threshold exceedence is indicated by a green LED on the front panel and the high threshold exceedence by a red LED When the trigger occurs the event is timestamped to the nearest second There...

Page 31: ...eshold actually creates two trip points one above the nominal and one below For example consider a setup where the nominal is 120 volts the low threshold is 6 and the high 12 The low trip points becom...

Page 32: ...be set as small as one second or as large as 255 seconds 5 12 Loose Neutral The Loose Neutral report shows whether the typical symptoms of a loose neutral have occurred This report is intended for si...

Page 33: ...required 12 volts specified by the range value of twice the nominal 240 volts If these voltages persist for 5 seconds in a row then a Loose Neutral record will be triggered If one voltage leg changes...

Page 34: ...an the size of four hours of Stripchart data at one minute intervals 5 13 1 Trigger Logic Waveform Capture uses a single threshold for triggering This threshold is a percentage At the end of each 60Hz...

Page 35: ...urbance can be seen in the Waveform Capture report The peak voltage length of the sag or swell and the coupling from phase to phase are easily seen in the graph Sometimes there are clues regarding the...

Page 36: ...he higher the level of current harmonics Frequently the neutral current looks much less sinusoidal than the line currents due to the fact that some harmonics don t cancel out in a three phase system e...

Page 37: ...g from any breach of warranty Equipment Return If any PMI product requires repair or is defective call PMI at 800 296 4120 before shipping the unit to PMI If the problem cannot be resolved over the ph...

Reviews: