STB50
Display
Description (entered values are factory settings)
20,0
°
C
Limit value
pre alarm
2,0
°
C
Hysteresis
pre alarm
Analog Output
0..20
> OFF
0,0
°
C
Start
analog output
100,0
°
C
End
analog output
Pre-alarm relay hysteresis
Change of the value in the range 1°C (0,1°C) ... 100°C (100,0°C).
Alarm value (switching point) of the pre-alarm relay
Change of the value in the corresponding measuring range.
Function of the analog output
OFF
Analog output is switched off
0..20
Analog output 0-20mA and/or 0-10V
4..20
Analog output 0-20mA and/or 0-10V
0..20/22
Analog output 0-20mA and/or 0-10V, in case of error 22mA/11V
4..20/22
Analog output 4-20mA and/or 2-10V, in case of error 22mA/11V
3.6/4..20
Analog output 4-20mA and/or 2-10V, in case of error 3.6mA/1.8V
Start value of the analog output
Change of the value in the range -200°C (-200.0°C) ... 2000°C
(2000.0°C).
End value of the analog output
Change of the value in the range -200°C (-200.0°C) ... 2000°C
(2000.0°C).