44
Serial command:
NE
Description: Enable or disable the E-24 error for E/B ratio.
Usage:
NE(0 or 1)<cr>. 0 enables, and is the default. 1 disables.
Example:
Send “NE1<cr>”. The E-24 error alarm will be disabled. When the emission to
beam ratio is not met, the E-24 error will not flash or function.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
NI
Description: Set the neutralizer filament current.
Usage:
NI(0.0-20.0)<cr>
Example:
Send “NI5.1<cr>”. The neutralizer filament current will bet set to 5.1 A. After the
source is turned on the power supply will ramp the neutralizer filament current to
5.1 A. Acceptable range is from 0.0 to 20.0 A. The source check-out sheet should
be consulted for normal ranges.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
NK
Description: Enable or disable lock-out of the front panel keypad. The keypad lock-out will
disable all switches on the front panel, and no alarms will sound.
Usage:
NK(0 or 1)<cr>. 1 enables the keypad Lockout, 0 disables the lockout, restoring
user access via keypad, as well as the alarm buzzer.
Example:
Send “NK1<cr>”. The keypad switches will be locked, allowing no input from the
front panel. The power supply must be run from an externally connected input
device, such as a deposition system or connected computer.
Response:
<lf><cr>OK<lf><cr><eot>
Serial command:
NL
Description: Set the neutralizer filament current limit.
Usage:
NL(0.0-20.0)<cr>
Example:
Send “NL8.5<cr>”. The neutralizer filament current limit will be set to 8.5 A. If
the neutralizer current exceeds 8.5 A, an alarm will trigger. Acceptable range is from
0.0 to 20.0 A. The source check-out sheet should be consulted for typical ranges of
neutralizer filament current. The limit should be set slightly higher than the
maximum expected neutralizer filament current.
Response:
<lf><cr>OK<lf><cr><eot>