Dose range min. / max. as indicated on system
user interface.
Accuracy of dosimetric indication (DAP, CTDI):
±40%
Linearity of radiation output
< 0.1
Exposure time
3D
Pulsed, effective 1.5 ‐ 36 s as indicated ±10%
Pan (SmartPan)
2.5 – 15.6 s as indicated ±10%
ProCeph
0.1 – 1.6 s as indicated ±10%
SID
3D / Pan (SmartPan)
700 mm
Ceph
1700 mm (66.9 in.)
Magnification
3D
1.40 ‐ 1.71
Pan (SmartPan)
1.40
Ceph
1.13
Duty cycle for height adjustment
25 s ON / 400 s OFF
Line voltage
100 ‐ 220 V~ / 50 ‐ 60 Hz
230 ‐ 240 V~ / 50 Hz
Line current
8 ‐ 17 A
Input power
Stand by
150 VA
Exposure
1800 W
Line harmonics
Cos better than 0.9
Max. permissible apparent impedance of supply
mains
0.5 ohm (100 VAC)
Max. continuous heat dissipation
250 W
Internal fuse(s)
One user replaceable fuse
100 ‐ 220 V~ / 16A FF H 500 V
230 ‐ 240 V~ / 8A FF H 500 V
Type
195100 ELU
External fuse(s)
100 ‐ 220 V ~ / 16A min. - 20A max. T 250 V
230 ‐ 240 V ~ / 10A min. - 20A max. T 250 V