TECHNICAL SPECIFICATIONS
12 Planmeca SingLED operating light
User’s Manual
9
TECHNICAL SPECIFICATIONS
Planmeca SingLED
Illumination intensity of 700 mm
8000 lx - 35000 lx (factory limitation)
Optimum colour temperature
5000-5500K
Classification
Class I, type B
Colour of the operating light arm
RAL 9016
Weight of SingLED light
8.1 kg (18.3 lbs)
Connection to supply
+15 VDC/0.9 A…+32 VDC/0.4 A
24 VAC/0.6 A50/60 Hz
Separation for main supply 4 kV
Air clearances 5 mm
Creepage distances 8 mm
Power supply must fulfill IEC60601-1 ed.3
(2005)
Power input
< 25 W
Power control (MENB1030A2402F01) (Optional)
Connection to supply
100 - 240 VAC
Supply frequency
50 / 60 Hz
Power output
24V/1.33A max
NOTE: When the Planmeca SingLED operating light is connected to other
equipment, the current must be limited to max. 2 A with e.g. a fuse.
Environmental requirements
Operating conditions:
Temperatures:
+15°C to +35°C (+59°F to +95°F)
Relative humidity:
Non-condensing humidity; 5% RH to 95% RH
Air pressure:
800 hPa to 1060 hPa (12 psi to 15 psi)
Altitude:
< 2000 m (less than 1.25 miles)
Storage conditions:
Temperatures:
-5°C to +60°C (+23°F to +140°F)
Relative humidity:
Non-condensing humidity; 5% RH to 95% RH
Air pressure:
700 hPa to 1060 hPa (10 psi to 15 psi)
If the unit has been stored at temperatures below +10 °C (50 °F) for more than few
hours, time must be allowed for the unit to reach the room temperature in its own
package, before connecting the dental unit to mains voltage.
Transportation conditions:
Temperatures:
-20°C to +60°C (-4°F to 140°F)
Relative humidity:
Non-condensing humidity; 5% RH to 95% RH
Air pressure:
700 hPa to 1060 hPa (10 psi to 15 psi)