Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
Planmeca Proline EC X-ray unit E-47
CEPHALOSTAT ADJUSTMENT
Technical Manual
Adjust the head support angle with the two screws located on the left side of the mechanism
(use 2.5mm allen key) (see Fig. 84) so that the sensor alignment tool is perpendicular to the
cephalostat head support.
Figure 84
Spirit level
90°