Chapter F - ADJUSTMENTS
Planmeca Compact i F-3
MECHANICAL ADJUSTMENTS
Technical Manual
1.3 Adjusting balance of the OP delivery arm
Remove the cable guide plate.
NOTE
When replacing the cable guide plate to its position, note that the distance
between the end of the guide and the end of the adjustment plate must be 0-
20mm (0-0.8 in.).
Adjusting the balance
The adjustment plate, which is connected to the gas spring, is held in position with the spring
support. To increase or decrease the tension of the gas spring, the position of the spring sup-
port compared to the adjustment plate must be altered. This is done with the aid of the adjust-
ing hook, which is pushed by a screwdriver (use a robust, round-shaped (not six-edged)
screwdriver with a diameter of 4.5…5.5 mm).
0...20 mm
Wireguide_4.2.eps
Adjustment plate
Cable
guide
plate
Adjustment plate
Gas spring
Spring support
Adjusting hook
y-spring
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......