Chapter F - ADJUSTMENTS
F-2 Planmeca Compact i
MECHANICAL ADJUSTMENTS
Technical Manual
1.2 Adjusting friction of the OP delivery arm joints
S
C
D
A
B
8 8
8 8
light
E
M
S
rotaf
.eps
To adjust the rotational friction of the console
arm column:
1. Remove the small cover plate from the cus-
pidor cover.
2. Unscrew the two M5x10 screws and
remove the holder plate.
3. Adjust the friction with the adjustment screw
located on the console arm column.
4. Reassemble in reverse order.
Adjust the rotational friction of
the console arm with the 4 mm
allen key. Adjust the two
screws equally to avoid wear-
ing. Tightening the screws
increases the friction.
Adjust the lifting friction of
the console arm with the 4
mm allen key. Tightening
the screw increases the
friction.
Adjust the rotational friction of the instrument
console with the 4 mm allen key. Adjust the
two screws equally to avoid wearing. Tight-
ening the screws increases the friction.
Summary of Contents for Compact I
Page 1: ...technical manual 10007462_23 EN...
Page 2: ......
Page 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Page 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Page 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Page 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Page 216: ......
Page 218: ......
Page 220: ......
Page 222: ......
Page 224: ......
Page 226: ......
Page 228: ......
Page 230: ......
Page 232: ......
Page 234: ......
Page 236: ......
Page 238: ......
Page 240: ......
Page 242: ......
Page 243: ......