background image

1.3  —  About the Invitium

optional battery pack

The Invitium Battery Pack is a system option that supplies a reg12 VDC to the
Invitium Medical Workstation.

The battery pack contains a sealed lead acid (SLA) battery, medically certified universal input
power supply, charging and monitoring electronics, and output voltage regulator. It is contained
within a plastic enclosure with all components mounted to an internal sheet metal frame to
provide mechanical strength as well as EMI/EMC protection and shielding.

The battery pack provides up to eight hours of system operation at a nominal load of 60 watts.
Recharge time is less than four hours.The charging and monitoring electronics control the
charge rate of the battery as well as the maximum discharge level.This is to ensure safe opera-
tion of the battery and to prevent damage from high charge rates and low discharge levels.

The Invitium system’s battery pack is certified by Underwriters Laboratories to medical stan-
dard UL2601-1 and C22.2 No. 601.1 M1990 for sale in the U.S. and Canada.The system is also
CE marked for use in the Europe.

For a full list of battery pack specifications, please refer to Section 5:Technical Specifications in this
manual.

The Invitium Medical Workstation is a Class I system under the MDD, intended for near-patient
proximity use. It is not intended for direct connection to the patient.

1.4  —  Safety Instructions

The Invitium Medical Workstation is designed to ensure both the highest level of product quality
and safety for the user.To maintain both quality and safety, follow the instructions in this manual
and these safety guidelines.

1.

Read the safety and installation guidelines carefully.

2.

Keep this manual handy for future reference.

3.

Install and use the Invitium system only on a sturdy surface and in stable surroundings.

4.

Do not place the Invitium workstation near a window. Exposing the system to rain, water,
moisture or constant direct sunlight can severely damage it.

5.

Do not place anything on top of the monitor-to-computer signal cord. Make sure the cord
is placed where it will not be stepped on.

6.

Do not apply excessive pressure to the monitor screen. Excessive pressure may cause per-
manent damage to the display.

7.

The Invitium workstation has no internal user-serviceable parts. Refer all servicing to
qualified personnel to maintain your warranty.

8.

Do not cover or obstruct any venting holes on the computer or the monitor.

9.

Store the Invitium system within –20 to +65 degrees Celsius. Storing the system outside
that temperature range could result in permanent damage.

10. If any cord or cable is frayed or damaged, immediately replace it with another of the same

type and rating as supplies by Planar.The safety and regulatory listing and certifications are
based on cables supplied by Planar.

4

Invitium

Medical Workstation

O P E R A T I O N S M A N U A L

Invitium Manual LAYOUT  3/26/02  9:35 AM  Page 4

Summary of Contents for invitium

Page 1: ...OPERATIONS MANUAL...

Page 2: ...Invitium Medical Workstation OPERATIONS MANUAL Copyright 2002 Planar Systems Inc...

Page 3: ...ay not be reproduced stored in a retrieval system or transmitted by any means without prior written permission of Planar Systems Inc Invitium is a trademark and Planar is a registered trademark of Pla...

Page 4: ...plaining its features and key specifica tions Section 2 Getting Started includes instructions on mounting the Invitium workstation and connecting its modules making display adjustments and calibrating...

Page 5: ...responsible for compliance could void the user s authority to operate the equipment Canadian Doc Notice For Class B Computing Devices This digital apparatus does not exceed the Class B limits for radi...

Page 6: ...upply alone does not represent a functional medical device Hence Planar configured a minimal operating system to exercise the display The resulting data is made available to interested parties The dat...

Page 7: ...em should assure that it is used in such an environment EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT GUIDANCE RF emissions The Invitium System uses RF energy only for its CISPR 11 Group 1 int...

Page 8: ...f a typical commercial or hospital environment Surge IEC 61000 4 5 1KV differential mode 2KV common mode 1KV 2KV Mains power quality should be that of a typical commercial or hospital environment If t...

Page 9: ...han the compliance level in each fre quency range b Interference may occur in the vicinity of equipment marked with the following symbol Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3Vrms 150...

Page 10: ...rmonized standards published in the official journal of the European Communities Indicates that the product has been tested and certified with respect to electric shock fire mechanical and other speci...

Page 11: ...turned to its original configuration before any warranty service will be performed by Planar No goods are to be returned to Planar without prior authorization Buyer will be responsible for packaging a...

Page 12: ...r person s rendering of technical advice assistance or services in connection with the buyer s order of the good fur nished hereunder The Buyer is not relaying on Planar s skill or judgment to select...

Page 13: ...xii Invitium Medical Workstation O P E R AT I O N S M A N U A L...

Page 14: ...the monitor to the computer 15 Starting up the Invitium workstation 16 Adjusting monitor controls 16 Using a floppy drive 17 Calibrating the Invitium Workstation Touchscreen 18 Non touchscreen systems...

Page 15: ...nformation messages 41 Resetting the Invitium system 41 Reinstalling Windows and Drivers 42 Windows 2000 installation procedure 44 Touchscreen driver installation 45 Technical Specifications Environme...

Page 16: ...om the box please contact us in the U S at 866 475 2627 or 503 748 1100 or in Finland at 358 9 42001 Optional equipment and parts will be shipped separately 1 1 About the Planar Invitium Medical Works...

Page 17: ...e external power supply or by an optional battery pack Each provides a regulated 12 volts as the main system voltage The computer and monitor can be mounted together to form a single integrated unit o...

Page 18: ...supply is equipped with an IEC320 appliance coupler The power adapter includes an attached power cable that terminates in the required monitor power connector An LED indicates when power is applied A...

Page 19: ...intended for direct connection to the patient 1 4 Safety Instructions The Invitium Medical Workstation is designed to ensure both the highest level of product quality and safety for the user To mainta...

Page 20: ...continues to operate normally while being cleaned in a fashion normal for a hospital environment This includes cleaning with a damp mildly soapy cloth Drip protection is provided in accordance with I...

Page 21: ...6 Invitium Medical Workstation O P E R AT I O N S M A N U A L...

Page 22: ...c All other trademarks are the properties of their respective owners Planar Systems Inc 1400 N W Compton Drive Beaverton OR 97006 USA Phone 866 475 2627 503 748 1100 Fax 503 748 1493 Planar Systems In...

Reviews: