The charging controller board has four levels of power consumption: charging, running, 
stand-by, and off.

Charging mode

The batteries are charged using a constant current/constant voltage 
algorithm. A constant current of 3.0 amps max is applied to each battery 
until the voltage on the battery reaches 14.4 V. The current exponentially 
decreases to maintain 14.4 V on the battery. The cooling fan is on during 
this cycle. The +12VDC regulator circuitry is powered by the power 
supply, and the communication system reports the charge status of 
the batteries to the host. 

 Running mode

The board consumes power at a rate proportional to the power 
consumed by the load during its running mode. The board proves to 
be 88% efficient when connected to a 4.5 amp load. 

Stand-by mode

The board consumes about 8.5 mA of current when in stand-by mode.
 A fully charged and new battery lasts about 3 months in the battery 
system before it needs to be replaced.

Off mode

The board consumes about 350 µA of current when in the stand-by 
mode. A fully charged and new battery lasts about 3 months in the 
battery system before it needs to be replaced. The battery continues 
to drain at a rate of 350 µA. Batteries that are not connected to the 
controller board have a shelf life of 1 month 97%, 3 months 91%, 
6 months 85%, at 68° F. 

Summary of Contents for BAT-24CM

Page 1: ...y System Locate Battery Housing Identify Connector Ports Install Battery Fuses USING THE BATTERY Charge Battery Set Up Power Management MAINTAINING THE BATTERY Replace Battery Troubleshooting APPENDIX...

Page 2: ...uct does not imply any recommendation or endorsement This document contains proprietary information protected by copyright No part of this manual may be reproduced by any mechanical electronic or othe...

Page 3: ...e Battery Troubleshooting APPENDIXES Battery Disposal Specifications iii Invitium Battery System BAT 24CM Contents Product Information iv About the Invitium Battery System 1 Locate the Battery Housing...

Page 4: ...le is frayed or damaged replace it immediately with another of the same type and rating as supplied by Planar The safety and regulatory listing and certifications are based on cables supplied by Plana...

Page 5: ...aning guidelines The Invitium battery system continues to operate normally while being cleaned in a fashion normal for a hospital environment Use a damp mildly soapy cloth to clean the exterior chassi...

Page 6: ...nst water ingress IPX1 Drip Proof Equipment At a nominal load of 50 watts the battery system provides up to 4 hours of system operation The charging and monitoring electronics control the charge rate...

Page 7: ...APPENDIXES Battery Disposal Specifications 2 Invitium Battery System BAT 24CM Locate the Battery Housing The Invitium battery system ships already installed on the Invitium cart The power cord USB ca...

Page 8: ...fications 3 Invitium Battery System BAT 24CM Identify the Connector Ports All external connections to the battery system are located on the rear of the battery enclosure The design of the enclosure mi...

Page 9: ...sal Specifications 4 Invitium Battery System BAT 24CM Install the Battery Fuses You need the two battery fuses and a screwdriver for this installation To install a battery fuse 1 Push the fuse in unti...

Page 10: ...rce for charging Charge time The charging controller board is capable of independently charging the batteries while powering the attached system This ensures that the charge rate of the batteries is t...

Page 11: ...urce and 2 powering the load When the battery has been discharged for a long period of time it reaches a point at which it needs to cut off power to its load The battery system communicates to the hos...

Page 12: ...et Up the Power Management System All communications occur through the USB serial interface These are displayed through the Windows Power Options Properties dialog box on the host computer To set up p...

Page 13: ...slider to the point at which you want the alarm to trigger Click the Alarm Action button to do one or more of the following select the type of notification set a computer action run a specific progra...

Page 14: ...y System BAT 24CM 4 Select the notification you want one of the two types or both Click OK Alarm action An alarm action to hibernate or shut down can be selected in the Power Options Properties menu A...

Page 15: ...ll Battery Fuses USING THE BATTERY Charge Battery Set Up Power Management MAINTAINING THE BATTERY Replace Battery Troubleshooting APPENDIXES Battery Disposal Specifications 10 Invitium Battery System...

Page 16: ...3 Unplug the power cord USB cable and output power cable 4 Remove the two battery fuses 5 Pull out the battery enclosure carefully and move it to a flat and stable work surface The unit is heavily we...

Page 17: ...y Disposal Specifications 12 Invitium Battery System BAT 24CM To replace the battery 1 Loosen the two screws and remove the retaining bracket 2 Detach the four battery terminal cable connectors and mo...

Page 18: ...Set Up Power Management MAINTAINING THE BATTERY Replace Battery Troubleshooting APPENDIXES Battery Disposal Specifications 13 Invitium Battery System BAT 24CM To reinstall the battery system into the...

Page 19: ...t acting size 5 mm x 20 mm Over temperature protection is activated Contact Planar No power supplied to Invitium system Output power cord not connected properly Make sure that output power cord is con...

Page 20: ...ding soil and groundwater supply It is a federal law to properly dispose of lead acid batteries once they can no longer be used Once a battery is purchased full liability and responsibility lies on th...

Page 21: ...power supply battery and charger unit packaged in a sheet metal enclosure Features Universal AC input Regulated 12V DC output Fast charging Intelligent communications USB host Cooling fan Externally a...

Page 22: ...tery Troubleshooting APPENDIXES Battery Disposal Specifications Battery requirements Battery type Sealed lead acid quantity 2 Yuasa NP12 12 or B and B EP12 12 Capacity 12 Ah each Output voltage range...

Page 23: ...ng mode The batteries have some level of charge and the battery system is not plugged into an AC power source During running mode the attached load to the battery system is operational The batteries d...

Page 24: ...s to maintain 14 4 V on the battery The cooling fan is on during this cycle The 12VDC regulator circuitry is powered by the power supply and the communication system reports the charge status of the b...

Page 25: ...ging battery 5 charging controller board 17 18 cleaning guidelines v compliance 24 components identifying 3 controller board charging 17 18 customer support 22 D discharge protection 6 discharging bat...

Page 26: ...t Planar will arrange for return and repair of the non functioning product 1 Contact Planar via the web at http www planar com support In North America call 866 PLANAR1 866 752 6271 In Europe call 358...

Page 27: ...ion or failure to follow instructions supplied with the product b Repair or attempted repair by anyone not authorized by Planar c Any damage of the product due to shipment d Removal or installation of...

Page 28: ...OPERTY CAUSED BY ANY DEFECTS IN THE PRODUCT DAMAGES BASED UPON INCONVENIENCE LOSS OF USE OF THE PRODUCT LOSS OF TIME LOSS OF PROFITS LOSS OF BUSINESS OPPORTUNITY LOSS OF GOODWILL INTERFERENCE WITH BUS...

Page 29: ...ference to radio or television reception which can be determined by turning the equipment on and off the user is encouraged to try to correct the interference by using one or more of the following mea...

Page 30: ...ted medical device safety standards and for the European market EC marked displays based on compliance with counsel directive 93 42 EEC commonly referred to as the Medical Device Directive MDD The fol...

Page 31: ...ive which establishes compliance to the basic medical safety standards However EMC compliance can only be accomplished in the configured medical device or system and is the responsibility of the devic...

Page 32: ...ct to electric shock fire mechanical and other specified hazards only in accordance with US 2601 1 and CAN CSA C22 2 No 601 1 and IEC 60601 1 for medical equipment If this mark appears with the indica...

Reviews: