1  INTRODUCTION  

 

The HS1 is a PZT actuated linear nanopositioning stage of exceptional resolution and 
stability.  Two  HS1

’s  can  be  stacked  to  form  a  high  performance  XY

  or  XYZ 

nanopositioning system. The HS1 comes complete with a position sensitive detector for 
closed loop operation.  
 
TRANSLATION (

μm) (X

)  

10, 30, 50, 70 or 100 µm  

VOLTAGE RANGE (V)  

-15V to +150V  

RESONANT FREQUENCY  
UNLOADED (Hz)  

7000Hz (10µm), 4000Hz (30/50/70µm) 
1500Hz (100µm)  

STIFFNESS (N/

μ

m)  

5.2N/µm (10µm), 3N/µm (30/50/70µm) 
1.1N/µm (100µm) 

MAXIMUM LOAD (hor/vert) (kg)  

1/0.5  

CABLE LENGTH (m)  

>1.5m  

CABLE CONNECTION  

DB-9  

 
 

1.1 Unpacking the HS1 

Before unpacking the HS1 read this entire operation manual, paying special attention to 
the following section on 

“Handling the 

HS1

. Check the contents of the package against 

the shipping list and notify PIEZOCONCEPT immediately if any items are missing.  

1.2 Handling the HS1.  

The HS1 is a high precision scientific instrument and therefore requires special handling 
in  order  to  ensure  proper  operation.  Mishandling  can  cause  permanent  damage  to  the 
nanopositioning  stage.  To  ensure  a  long  and  useful  life  the  following  guidelines  should 
be strictly followed.  
 

 

Never  insert  anything  into  the  EDM  grooves

.  The  EDM  grooves  are  the  cuts 

that form the flexure hinges and separate the moving portion of the stage  

  Do not move the translation stage by pushing on it with your hands or any other 

object.  

  Avoid applying a torque between the moving stage and the frame.  

  Do not drop, treat roughly, or physically shock the Nanopositioning stage.  

  Do not lift by the cable.  

  The surface to which the  HS1 is mounted to should be flat and clean. Likewise, 

the bottom of the HS1 should be free of particles and dust before mounting.  

  Do not immerse in any liquid. If the HS1 requires cleaning slightly dampen a lint 

free cloth with iso-propanol or ethanol and lightly wipe the surface. Do not get any 
liquid or lint into the EDM grooves.  

  Never  disassemble  the  nanopositioning  stage,  there  are  no  serviceable  parts 

inside.  

 

Summary of Contents for HS1

Page 1: ...1 INSTALLATION AND OPERATION MANUAL HS1 PIEZOCONCEPT 15 Rue du Bocage 69008 LYON France ...

Page 2: ...om the PIEZOCONCEPT controller first set the command voltage to 0 0V then turn the AC power to the PIEZOCONCEPT controller off and finally wait one minute before disconnecting The HS1 has no user serviceable parts Only trained service personnel should perform service IMPORTANT All Technical Information recommendations and examples related to PIEZOCONCEPT Products made in this manual are based on i...

Page 3: ...ATION 5 2 1 Installing a single axis HS1 unit 5 2 2 Installing a two axis HS1 unit 6 2 3 Installing a three axis HS1 unit 6 3 GROUND LOOPS 7 3 1 Prevention and identification of ground loops 7 4 OPERATING THE NANOPOSITIONING STAGE 7 4 1 Operating in closed loop mode 7 4 2 Care during operation 8 ...

Page 4: ...cientific instrument and therefore requires special handling in order to ensure proper operation Mishandling can cause permanent damage to the nanopositioning stage To ensure a long and useful life the following guidelines should be strictly followed Never insert anything into the EDM grooves The EDM grooves are the cuts that form the flexure hinges and separate the moving portion of the stage Do ...

Page 5: ... 2 1 Installing a single axis HS1 unit To install a single axis HS1 use the following steps On the surface that the HS1 is to be mounted on tap M3 holes on a 52 50X52 50mm square Make sure the surface to be mounted to is clean flat and free of burs Using a lint free cloth gently wipe off the bottom of the HS1 to remove any particles or dust Use M3 Socket screws to secure the stage Use a maximum to...

Page 6: ...t head screws Do not hold on to the bottom or Y axis while loosening To install a two axis HS1 system use the following steps 1 Install the first or bottom HS1 Y axis using the procedures described above in Section 2 1 2 Using a lint free cloth gently wipe off the top of the installed HS1 and the bottom of the adapter plate to remove any particles or dust 3 Using four M2 5 screws provided bolt the...

Page 7: ...ews and the mounting surface may short circuit the ground loop In a few cases this may not be an effective method When this occurs please identify high current sources returning to ground through your mounting surface Mounting surfaces should never be used as the electrical ground current path for any instrumentation such as vacuum pumps computers etc Should you observe unexpected oscillations in ...

Page 8: ...ow 1 minute for the PZT actuators to discharge before disconnecting For more information see the PIEZOCONCEPT CONTROLLER OPERATION MANUAL 4 2 Care during operation The HS1 is a high precision scientific instrument and should be handled with care during operation Failure to do so may result in permanent damage During operation ensure that there are no physical constraints on the moving stage or any...

Reviews: