EZ-Check™ Switch/EZ-Step™ Push-button

For Source - Slide the switch to select from three user 
stored values for the desired calibration points. The 
HI, DIAL and LO values can easily be changed to suit 
the calibration requirements. Press the push-button 
like a stopwatch to start or stop the totalizer output.
For Read - Slide the switch to the DIAL position. The 
Model 541 will begin making readings.  Slide the 
switch to high and you will get the highest point read 
and then slide the switch to the low position and you 
will get the lowest range. Press the push-button like a 
stopwatch to start or stop the totalizer counting 
pulses.

w

 

SOURCE/AMPLITUDE Key

Press and release the SOURCE/AMPLITUDE Key to 
step through the source ranges.

Source CPH 1 - 20000
Source CPM 0.1 – 2000.0
Source HZ 0.01 – 200.00
Source HZ 0.1 – 2000.0
Source KHZ 0.001 – 20.000

 Totalizer

Press and hold SOURCE/AMPLITUDE key to enable 
amplitude adjustment. Turn the EZ-Dial knob 

u

  to 

adjust the amplitude voltage from 0.1-12.0 Vp.
Press and release the SOURCE/AMPLITUDE or 
STORE/CLEAR button to save your selection.

e

 

ON / OFF Key

Press the ON/OFF key to turn the Model 541 on or off.

Basic Operation

SOURCE

CPH  1 - 20000

20000

HI

r

 

MENU Key

Press and release the menu key to enable the 
configuration menu. This allows you to choose 
the settings for AUTO OFF, X1/X10, Zero Based/
Zero Crossing, EZ-Check HI/LO Readings, 
SINE/SQUARE Wave output and reset to BASIC 
CONFIGURATION.

t

 

READ/GATE TIME Key

Press and release READ/GATE TIME key to step 
through the read ranges.

Read CPH 1 - 20000
Read CPM 0.1 – 2000.0
Read HZ 0.01 – 200.00
Read HZ 0.1 – 2000.0
Read KHZ 0.001 – 20.000

 Totalizer

Press and hold the READ/GATE TIME key to 
enable time adjustment for Source & Read 
Totalize mode.
Turn the EZ-Dial knob 

u

  to adjust the totalize time 

from 1 to 100 minutes.
Press and release the SOURCE/AMPLITUDE or 
STORE/CLEAR button to save your selection.

y

 

STORE/CLEAR Key

For Source - press the STORE/CLEAR key to 
save the calibration values. The display will flash 
“STORED” to confirm.
For Read - press STORE/CLEAR to clear the 
values saved in the EZ-Check™ HI and LO 
positions. The display will flash “CLEARED” to 
confirm.

u

 

EZ-Dial™ Knob

For Source - turn the EZ-Dial knob to change 
the output frequency in increments of one least 
significant digit. Push down and turn for faster 
dialing with increments  of 100 digits.
For Read - turn the EZ-Dial knob to adjust the 
trigger level.

CHANGING BATTERY

Low battery is indicated by 

“BAT”

 on the display. 

Approximately one to four hours of typical 
operation remain before the 541 will automatically 
turn off. To change the battery remove the battery 
cover from the back of the unit. Replace with a 
single 9V battery. 

Note:

 Alkaline battery is supplied and 

recommended for maximum battery life & 
performance.

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Summary of Contents for PIECAL 541

Page 1: ...ptical pickups with the GATE TIME LED that is synchronized with the output pulses Match your instrument signal levels Choose zero based or zero crossing square or sine waves output from 0 1 V to 12 V peak to peak Measure from 0 1 V to 120 V peak with X1 or X10 attenuation Turn the dial to adjust the output amplitude to the level required Read signals from flow meters Measure flow meter vibration p...

Page 2: ...ero Based Zero Crossing EZ Check HI LO Readings SINE SQUARE Wave output and reset to BASIC CONFIGURATION t READ GATE TIME Key Press and release READ GATE TIME key to step through the read ranges Read CPH 1 20000 Read CPM 0 1 2000 0 Read HZ 0 01 200 00 Read HZ 0 1 2000 0 Read KHZ 0 001 20 000 Totalizer Press and hold the READ GATE TIME key to enable time adjustment for Source Read Totalize mode Tur...

Page 3: ...aks that are centered around zero Select zero based when the instruments you are calibrating require only positive signals Note To simulate negative only signals select zero based and connect the red lead to the ground and the black lead to the input connection of the instrument being calibrated Press and release the MENU or STORE CLEAR Key to save your selection EZ CHECK pressing the knob will to...

Page 4: ...y Instruments SOURCE CPH 1 20000 20000 HI Storing EZ Check Outputs STORING HI and LO EZ Check Outputs 1 Store your high SPAN output frequency by moving the EZ Check switch q to the HI position and turn the EZ Dial knob u until the desired frequency is on the display Press the STORE CLEAR Key and STORED appears on the display 2 Store your low ZERO output frequency by moving the EZ Check switch q to...

Page 5: ... 541 to the input of the device to be calibrated 5 Press the EZ Step push button q like a stop watch to start outputting pulses The word RUN will appear on the display and the 541 display increments with each pulse while the totalizer is running When the selected number of seconds has elapsed the word STOP appears on the display and the total number of pulses sourced during the time appears on the...

Page 6: ... adjust the totalize time from 1 to 100 minutes Press and release the READ GATE TIME or STORE CLEAR button to save your selection 3 Connect the 541 to the output of the device to be calibrated 4 Turn the EZ Dial knob u to adjust the trigger level of the 541 The LED will light when the 541 locks onto the input signal indicating that it is properly detecting the signal 5 Press the EZ Step push butto...

Page 7: ...Rejection 50 60 Hz 50 dB 50 60 Hz 90 dB Read Frequency Input Impedance 1 Megohms 100pF Signal Attenuation Adjustment Range Adjustable trigger level with X1 0 1 V peak to 12 0 V peak and X10 1 V Peak to 120 V peak Note on reading slow sine waves High signal noise and low slew rate Volts per second will affect reading uncertainty Source Frequency Output Impedance 25 Ohms Source Current 6 mA at 12 V ...

Page 8: ...warranty can be made by returning the equipment prepaid to our factory The equipment will be repaired replaced or adjusted at our option The liability of Practical Instrument Electronics PIE is restricted to that given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electro...

Reviews: